Prima C/ Prima CS COLPOSCOPE 

  INTRODUCTION (continued)

6128000-795 

 Prima C/ Prima CS 

Issue 2.2
Printed on November,2020

2

IMPORTANT GENERAL REFERENCES 

Preliminary notes

LABOMED  is  renowned  for  providing  innovative  technologies  and  products  in  women’s  healthcare.  PRIMA
C  /  Prima  CS  colposcopes  were  developed  in  cooperation  with  practicing  gynecologists.  Experience,
precision,  and  continuous  manufacturing  optimization  has  enabled  LABOMED  to  became  the  leading
Manufacturer  of colposcopic precision instruments, with the largest product range of colposcopes in the
world.

Determination-Appropriate Use

Only LABOMED accessories, bases, and source of light may be utilized with LABOMED colposcopes.

Restrictions, which are stipulated in the directions for use, are to be considered.

The device is designed for application in medical practices and/or clinics.

According  to  the  stipulation  of  the  MPG (German Medical Device law), colposcopes may be operated
only by those persons who have been instructed in the expert handling of the device.

The proper installation and maintenance of the device are vital prerequisites before using the colposcope.

The user should check the functional stability and proper condition of the device before every procedure.

Indications for use

A colposcope may be helpful in the following situation:
1. Major abnormality on pap smear
2. Persistent minimal abnormality on pap smear
3. Abnormalities of squamous or columnar cells
4. Persistence of an inflammatory appearance of the smear despite adequate treatment
5. Presence of keratinized cells suggestive of leukoplasia
6. Symptoms of metrorrhagia, in particular contact bleeding
7. Abnormalities appearance of cervix or vagina by naked eye examination
8. Evaluation after exposure to diethylstilbestrol(DES) or related drugs during embryo genesis
9. Monitoring after treatment of dysplasia, regardless of treatment method
10. Evaluation of anogenital condylomatosis
11. Evaluation of vulva-perinea bowenoid papulosis
12. Preoperative evaluation before genital-tract surgery: hysterectomy, Treatment of genital prolapse,cervico
       vaginal plastic surgery, etc.
13. Lesion noted on routine gynecologic examination

Содержание Prima C

Страница 1: ...S Federal Law restricts this device to sale by or on the order of a licensed Physician Rx only To ensure proper use of this instrument as well as to avoid injury while operating Instrument understandi...

Страница 2: ...ma C Operating Instructions Prima C 9 1 On Off Switch 9 2 Brightness control Control elements Prima C CS Instructions for using Microscope a Adjustment of Interpupillary Distance b Changing the magnif...

Страница 3: ...ABOMED dealer or from the LABOMED service department Contact information is provided at the end of this guide LABOMED Prima C CS COLPOSCOPE is an optical instrument used in a medical diagnostic proced...

Страница 4: ...ructed in the expert handling of the device The proper installation and maintenance of the device are vital prerequisites before using the colposcope The user should check the functional stability and...

Страница 5: ...hese operating instructions These instructions are structured around the operation of the colposcope from assembly to the finished system Basic troubleshooting information is also provided Our goal is...

Страница 6: ...ification Knob Tilt Locking Knob CMO Focus Knob Eyepieces Binocular Head Binocular Head Locking Knob Arm Locking Knob Joystick Intensity Control Knob Prima C Ergonomic stand for easy and accurate posi...

Страница 7: ...CMO Fine Focusing Knob Magnification Knob Tilt Locking knob Suspension arm hydraulic movement knob Suspension arm spring Tension Adjustment Prima CS Colposcope can be positioned easily with the help...

Страница 8: ...D APPROPRIATE INSPECTION AND TESTING MUST BE CONDUCTED TO ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT WARNING ENSURE THAT THE VOLTAGE APPLIED TO THE UNIT IS THE SAME AS THE VOLTAGE THAT IS INDICATED...

Страница 9: ...CE SECTION FOR DETAILED CLEANING INSTRUCTIONS CAUTION MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS WITH RESPECT TO ELECTROMAGNETIC CHARGE EMC AND NEEDS TO BE INSTALLED AND SERVICED ACCORDING...

Страница 10: ...way from rain Connection ON OFF Year of manufacture used on PRODUCT DATA PLATE Alternating Current Power Fragile content of the transport package are fragile and therefore shall be handled with care C...

Страница 11: ...Optics and Optical instruments Medical endoscopes and endoscopic accessories Part 3 Determination of field of view and direction of view of endoscopes with optics ISO 8600 5 First edition 2005 03 15 O...

Страница 12: ...power supply Refer Fig 2 1 Page 13 Microscope Arm Refer Fig 2 2 Page 13 Microscope Head with Eyepieces Inclined or Straight Refer Fig 2 13 Page 21 Fasteners are enclosed in the packing box PRIMA CS T...

Страница 13: ...TING OF ROLLER STAND AND COLUMN Fig 1 Insert column 1 into the mobile base 2 Fasten screw 3 with enclosed hexagon socket wrench to column 1 The mobile bases have 5 rollers 4 out of two have locking br...

Страница 14: ...ation box 3 to the column against the vertical guide 2 Screws 1 to be tighten to hold the box as shown Fig 2 1 Insert the arm 4 of the microscope head on the top of the column 5 Screws 6 to be tighten...

Страница 15: ...provided from the bottom of metal plate Refer Fig 2 3 Install the column D on the base Engage the column to the indexing screw E Align the tapped hole of the Column in the screw seat hold the column...

Страница 16: ...screw driver in the socket situated in Suspension arm front coupling as shown as 7 in fig 2 5 The same is hanged with a tie to Potentiometer knob 5 1 SPECIAL INSTRUCTION The joint which connects swive...

Страница 17: ...olposcope with H Base Stand Total weight Approx 86 5kg Three Boxes containing Base left wheel support 20 kg Base right wheel support 20 kg Base center Support 20Kg One Box Containing Column 8Kg Micros...

Страница 18: ...w Instruction on electrical label provided at bottom of the arm as shown 3 here in Fig i 8 1 REPLACING THE ILLUMINATION SOURCE Open the swivel arm cover s Detach the fibre optic cable and replace the...

Страница 19: ...Prima C Prima CS COLPOSCOPE 17 6128000 795 Prima C Prima CS Issue 2 2 Printed onNovember 2020 8 3 WIRING CODING DIAGRAM...

Страница 20: ...as shown in figure number 2 8 below a CHANGING OF FUSE Note Fuses are in fuse holder F 2 8 2 main input fuses for lamp supply for mains voltage 100 240 V Changing of fuse Draw out the plug Insert a sc...

Страница 21: ...n box After the appliance is swtiched on the green illumination switch diode indicates that the microscope is ready for operation The LED starts burning and the cooling fan starts working 9 2 BRIGHTNE...

Страница 22: ...knob to change the Magnification 1 Common Main Objective CMO 2 Fine focusing knob 3 Joystick to tilt the head arm forward and backward 4 Head arm locking knob 5 Knob to lock the arm 6 Knob to swing in...

Страница 23: ...ic movement knob 11 Magnification Knob 19 Suspension arm spring Tension Adjustment 12 Fine Focusing Knob 20 Microscope arm locking knob 13 Head Locking Knob 21 Suspension arm locking knob 14 Swivel ar...

Страница 24: ...Fig 2 13 Turn the microscope in working position Inclined tube 45 1 Adjust eyepieces 2 to the required IPD by turning the lateral knob 3 Straight tube 4 Adjust eyepieces 5 by moving the binocular tube...

Страница 25: ...N Fig 2 14 Adjust to the highest magnification with any of the rotating knobs 1 on the magnification changer See to it that the magnification step changer is engaged into its index position related to...

Страница 26: ...mplete microscope in the focus range by any of the following means a Roll the mobile stand forward and backward b Adjust the arm towards left right and up down c Adjust the microscope angle left and r...

Страница 27: ...ON FIG 2 16 Swing filter knob 1 to swing in or swing out the red free Green filter This will help in filtering the red tissue in the image being observed and will highlight the enhance vessel image an...

Страница 28: ...NS FOR USING THE MICROSCOPE continued e EYEPIECES WITH DIOPTER Fig 2 17 You can adjust the diopter settings as per your vision correction Diopter adjustment is 5D Fig 2 17 26 6128000 795 Prima C Prima...

Страница 29: ...Fig 2 18 The objective 1 has screw type mounting Unlock it by turning anti clock wise and lock by turning clock wise Eyepieces 2 are insert able Pull out for changing Other objectives eyepieces can be...

Страница 30: ...straight or 45 inclined 2 Total four accessories are provided as follows 2 1 DSLR camera adapter for Nikon Canon and Sony 2 2 CCD camera adapter for Mintron 2 3 Adapter for Sony Handy Cam 2 4 Assisto...

Страница 31: ...unscrew optical system and Mechanical parts unless explicitly instructed to do so in this manual Protect the microscope from oil and grease Never oil or grease the guide surfaces or mechanical parts T...

Страница 32: ...d store them in a clean environment only Store under UV light when not in use Use in continuously climate controlled rooms only Keep moisture away using silica gel and cover with a plastic cover OCCUP...

Страница 33: ...st it to feel wet Disinfect clean the surface gently Note 1 Use mask and Gloves while performing cleaning and disinfection 2 While cleaning muslin cloth should not be dripped wet to prevent seepage an...

Страница 34: ...A C Prima C Prima CS COLPOSCOPE Schedule of Autoclavable caps Prima C Autoclavable caps C1 Part no 6122015 209 C2 Part no 6168000 219 C3 C4 Part no 6168000 213 32 6128000 795 Prima C Prima CS Issue 2...

Страница 35: ...Autoclavable caps Prima CS Autoclavable caps C5 C6 C8 Part no 6122015 209 C7 Part no 6168000 219 C9 C10 C11 C12 Part no 6168000 213 Prima C Prima CS COLPOSCOPE 33 6128000 795 Prima C Prima CS Issue 2...

Страница 36: ...s Prima C Prima CS COLPOSCOPE 34 6128000 795 Prima C Prima CS Issue 2 2 Printed onNovember 2020 For Operation For transportation and storage Temperature Rel humidity without condensation Air Pressure...

Страница 37: ...Use constant voltage stabilizer to avoid sudden input voltage variation and to enhance LED life PRIMA CS H Base Stand Column lockable wheels Straight Observation tube interpupillary distance 50 75mm O...

Страница 38: ...hone 510 445 1257 Fax 510 991 9862 Email sales laboamerica com www laboamerica com EU REP A16372 LABOMED ISO 13485 Labomed Europe Essebaan 50 NL 2908 LK Capelle a d IJssel The Netherlands Tel 31 0 10...