Prima C/ Prima CS COLPOSCOPE
CARE & MAINTENANCE (continued)
30
6128000-795
Prima C/ Prima CS
Issue 2.2
Printed on November,2020
• Protect the microscope from oil and grease. Never oil or grease the guide surfaces or mechanical
parts.
• Remove coarse contamination using a damp disposable cloth.
• Use disinfectants based on the following ingredients: aldehyde, alcohols, quaternary ammonium
compounds.
•
Camera: Clean optical components using a lint-free cloth. Soak the cloth using a little methanol or
glass cleaner. Do not use ethanol and spirit.
•
Do not clean products and optical components in a cleaning/disinfecting device or ultra sound bath.
•
LABOMED Maxi Lite coatings are fungal resistant. If you clean as described above, the coatings will
not be damaged.
TROPICAL ENVIRONMENT/FUNGUS:
LABOMED employs certain safety precaution in its manufacturing techniques and materials. other
preventive measures include:
•
Keep optical parts clean.
•
Use and store them in a clean environment only.
•
Store under UV light when not in use.
•
Use in continuously climate-controlled rooms only.
•
Keep moisture away using silica gel and cover with a plastic cover.
OCCUPATIONAL SAFETY AND HEALTH PROTECTION:
Observe work safety and health protection of persons responsible for processing contaminated products.
Current regulation of hospital hygiene and prevention of infection must be observed in the prepara-
tion, cleaning and disinfection of the products.
INSTRUCTIONS
WORKPLACE:
Remove surface contamination with a paper towel.
REPROCESSING:
Recommended reprocess a product immediately after use or as & when required as per below
cleaning instructions.
CLEANING & SERVICING:
Needed: water, detergent, spirit, microfiber cloth
• Take a Linen or any soft cloth. Moist it Slightly with running tap water (<40°C), using a little detergent
and clean the metallic and plastic parts.
• Clean all optical components with spirits or alcohol.
• Dry optical components using a microfiber cloth; dry the rest of the product using a paper towel.
• For servicing as and when required, inform LABOMED after- sales service department.