Replacing the fuse
Before replacing a fuse, it is recommended to find the probable
causes which caused it to blow.
We also recommend disconnecting all cables before opening the
box.
To replace the fuse, proceed as follows:
1.
Get a 4 Amp, 5x20mm fast-acting fuse.
If the scale is supplied with the
OUT
connector, get an 6
Amp, 5x20mm fast-acting fuse.
2.
Using a Phillips N. 1 screwdriver, remove the 13 fixing screws
from the rear cover.
3.
Open the rear cover.
4.
Identify the fuse holder on the circuit board (see photo 2).
5.
Remove the fuse holder cap and replace the fuse.
6.
Put the cap back on to the fuse holder and close the box
making sure not to crush the connection wires.
7.
Screw the 13 fixing screws back onto the rear cover.
Photo 2