Chapter 2: Prerequisites
Chapter 3: Getting Started
P#o%&ct Se#+ice -./00.122.341/
28
Tubing or pipe f
"
P1.3 cmQ ID should be used for the
drain.
DO NOT ROUTE THE DRAIN TO AN
ELEVATION MORE THAN 30
"
P76 cmQ
ABOVE THE PUMP DISCHAR5E. DO
NOT REDUCE THE SIgE OF THE
DRAIN PLUMBIN5.
DO NOT REMOVE THE HOSE FROM
THE CLAMP ON THE REAR OF THE
5LASSWARE WASHER.
The drain piping can be routed to either a sink or a floor
drain. If you route the drain piping to a sink drain without
using an air gap, the highest point of the drain pipe must
be above the sink drain, as shown in Figure 3-13.
☞
The use of an air
gap is strongly
recommended for
all installations to
prevent the siphon-
ing of waste water
into the washer.
If an air gap is not
used, the drain line
must be elevated
above the water
level of the sink.
Figure 3-13
☞
Содержание 44000 Series
Страница 14: ...Chapter 2 Prerequisites Chapter 2 Prerequisites P o ct Se ice 00 122 341 10...
Страница 54: ...Chapter 2 Prerequisites Chapter 4 Using Your Washer P o ct Se ice 00 122 341 50...
Страница 74: ...Chapter 2 Prerequisites Chapter 6 Modifying Your Washer P o ct Se ice 00 122 341 70...
Страница 79: ...Chapter 7 Troubleshooting 75 P o ct Se ice 00 122 341...
Страница 81: ...Appendix A FlaskScrubber and SteamScrubber Components 77 P o ct Se ice 00 122 341 Figure A 1...
Страница 83: ...Appendix A FlaskScrubber and SteamScrubber Components 79 P o ct Se ice 00 122 341 Figure A 2...
Страница 90: ...Chapter 2 Prerequisites Appendix B Glassware Washer Dimensions P o ct Se ice 00 122 341 86...
Страница 91: ...Appendix B Glassware Washer Dimensions 87 P o ct Se ice 00 122 341...
Страница 97: ...Appendix C Glassware Washer Specifications 93 P o ct Se ice 00 122 341 Wiring Diagram 230V Models...
Страница 107: ...P o ct Se ice 00 122 341 103...