9-4 TROUBLE SHOOTING
The DPM-900/950 series provides several error indications.
The error indication, status at that time, and remedies are described in the following table. If any error
indication appears on the monitor, check the following table for remedies.
Error
Displayed Timing
Status
Remedy
After doing bridge
check
or
After the setting the
internal gain
(ACTL LOAD mode)
1) The sensor is not connected.
1) Connect the sensor.
2) The bridge box is not
connected or the strain gage
(s) is not connected to the
bridge box.
2) Connect the bridge box or strain
gage.
3) Cable is broken.
3) Repair or replace the cable.
4) Input strain exceeds ±10000
µm/m.
4) Make input strain within ±10000
µm/m.
After doing bridge
check
Not applicable sensor is
connected.
Connect the applicable sensor.
After turning ON the
power, when the
input from the
external OSC is not
coming (slave only)
OSC frequency or voltage is
faulty.
Replace the normal master.
After setting the
internal gain
(normal mode)
Less than minimum internal
gain (×200) is set.
Decrease the setting value of the
CAL switch or increase the setting
value of the VOLTAGE OUT
switch and set the internal gain
again.
After setting the
internal gain
(normal mode)
Higher than maximum internal
gain (×20000) is set.
Increase the setting value of the
CAL switch or decrease the setting
value of the VOLTAGE OUT
switch and set the internal gain
again.
After setting the
internal gain
(TEDS mode)
1) TEDS-installed sensor is not
connected.
1) Connect the TEDS-installed
sensor.
2) TEDS information cannot be
read. (Cable is broken or the
sensor is faulty.)
2) Repair the cable or replace the
sensor.
After setting the
internal gain
(TEDS mode)
The applied voltage is outside
the range of the sensor's
recommended excitation
voltage.
From the rear setting switch, set the
bridge voltage to the appropriate
value and reboot.
Or replace the sensor that the
sensor’s recommended excitation
voltage matches the current bridge
voltage.
After setting the
internal gain
(TEDS mode)
Less than Minimum internal
gain (×200) is set.
Increase the setting value of the
VOLTAGE OUT switch and set the
internal gain again.
After setting the
internal gain
(TEDS mode)
1) Higher than maximum internal
gain (×20000) is set.
2) Setting value of VOLTAGE
OUT switch exceeds 10.00 V
1) Decrease the setting value of the
VOLTAGE OUT switch and set
the internal gain again.
2) Setting value of VOLTAGE OUT
switch set 10.00V or less.
After setting the
internal gain
(ACTL LOAD mode)
1) Less than minimum internal
gain (×200) is set.
2) Input strain is too large.
3) Input strain is minus.
1) Increase the setting value of the
VOLTAGE OUT switch and set
the internal gain again.
2) Reduce the input strain.
3) Set the input strain to plus.
After setting the
internal gain
(ACTL LOAD mode)
Higher than maximum internal
gain (×20000) is set.
Decrease the setting value of the
VOLTAGE OUT switch and set the
internal gain again.
Contact KYOWA or our representatives even if the error indication appears after conducting the remedies.