43
9. APPENDIXES
9-1 BRIDGE CHECK FUNCTION AND RESULT INDICATION
The following table shows the relation between error indications on the monitor and detected
wire-breaking locations.
When 1 wire is broken, the monitor indicates “wire-breaking location” exactly. But if 2 wires are broken,
wire-breaking locations of only one wire are shown as described in the following table. In case 3 or 4 wires
are broken, appears all the time.
In addition, if a side of the bridge is broken, an error indication does not necessarily agree with its location.
If appears, it indicated that the bridge check circuit is faulty. In this case, contact
KYOWA or our representatives.
NOTE
Wire-breaking locations
Error
None