33
6-1-3 Wiring to Bridge Box
How to configure a variety of strain gage bridges is described below.
6-1-4 Shield Conversion Cable N-117
When using a bridge box that can input multiple channels
with DPM-911/912/913, please use shield conversion cable
N-117 (optional).
It can be isolated for each channels.
When wiring strain gage (s) to the bridge box, soldering is recommended to reduce contact
resistance.
Use shortest possible leadwires for connecting the strain gage and the bridge box. Leadwires
shorter than 1 m are ideal.
Use a shielded cable for dynamic strain measurement. It is advantageous to reduce noise.
For other details, refer to the bridge box’s Instruction Manual.
NOTE
Gage Method
Bridge Circuit
Wiring to the bridge box
Quarter bridge system
(2-wire)
Quarter bridge system
(3-wire)
Half bridge system
Half bridge system
(3-wire active-active)
Full bridge system