29
QuickGene-Mini8L Operation Manual
3 Maintenance
3.2 Cleaning the Parts
The following table describes how to clean the Parts of QuickGene-Mini8L.
Parts
Cleaning methods
Cartridge Holder
Clean the parts with neutral detergent then wash out with 0.1% sodium
hypochlorite solution. Next, wash the solution out with DEPC water and
then dry the parts.
Tube Holder
Pressure Seal Plate
3.3 How to replace the Pressure Adapter
(1) Turn off the power.
Disconnect the plug of the power cable from the outlet.
Plug
(2) Pull out the Cartridge Holder and Tube Holder.
(3) Hold the pressing type knobs of both sides of the
Pressure Adapter as shown in the figure, and then
press the upper knobs in the direction of the arrow.
Knobs
3