KRUUSE 271390 Скачать руководство пользователя страница 20

VI. Cleaning and Sterilization 

In this section we describe the procedures to clean and sterilize the scaler.    It is 
important to follow these procedures before using the machine, otherwise, patients 
and/or doctors may have the possibility of getting infection.     

It is mandatory 

that doctors wear sterilized gloves during these procedures at all time to avoid 
any possibilities of incomplete sterilization and/or infection.   

Below we detail 

the infection control procedures for the Handpiece, the Tip, and the Main Unit, 
respectively. 

 

Handpiece   

Before cleaning, remove the Tip from the Handpiece.    Let the Handpiece run 
for a couple of seconds to drain out the water and any possible contamination 
left inside the Handpiece. The outer surface of the handpiece should be 
cleaned with anantiseptic soap or solution. Rinsed off with water and wiped or 
sprayed with a chemical disinfectant that is compatible with the handpiece 
material, such as a 1:6 dilute solution of Sporicidin brand disinfectant. A 
sterile insert or TIP is then reassembled to the handpiece in preparation for the 
next patient.   

Warning

Do not put the Handpiece and the extension wire directly into 

the sterilized fluid.  Any left such fluid inside the machine 
will interfere with the normal operations of the system.

 

At the end of the day with the Tip remove, the handpiece and cable should be 
scrubbed with an antiseptic soap or solution and rinsed off with water. The 
handpiece should be scrubbed a second time with an antiseptic soap or 
solution and rinsed off with water. 
 

Warning:

 

The chemical disinfectant should not be allowed to remain on 

the surface longer than the recommended time or material 
damage may result.

 

Note

Cleaning of the Hand-piece is suggested after each patient use.

 

 

Tip 
 

After each usage, there will be saliva and/or blood or other debris left on the 
Tip, consequently, it is necessary to clean the Tip with a supersonic cleaner 
first. This can be done manually by scrubbing with a brush or by use of an 
ultrasonic cleaner with a solution of detergent and water. After scrubbing. The 
Tip should be rinsed thoroughly with water to remove all detergent. Then dry 
the rinsed Tip. Finally, put the Tip in a bag and then put it into a medical-
equipment autoclave and sterilize the Tips at 260

(127

) for 30 minutes or 

as recommended by the manufacturer of the particular sterilizer used. 
An alternative, but less desirable method of sterilization is the use of a 
compatible chemical sterilant such as Sporicidin brand disinfectant. In this 
method, the Tips is immersed at room conditions in the disinfectant at the 
appropriate strength and for the recommended period by the manufacturer. The 
thoroughly rinsed with sterile water and completely dried. 

 

 

 

14

Содержание 271390

Страница 1: ...Instruction Manual KRUUSE ART SP2 Piezo scaler and micromotor Cat No 271390 J RGEN KRUUSE A S Havretoften 4 DK 5550 Langeskov Tlf 45 7214 1511 Fax 45 7214 1400 www kruuse com I...

Страница 2: ...ends to enforce its rights to this manual under copyright law No one may copy the materials presented in this manual without the express written permission of BONART MEDICAL TECHNOLOGY INC Product spe...

Страница 3: ...MENTS NOTE Indicates some tips CAUTION Indicates correct operating or maintenance procedures in order to prevent damage to or destruction of the equipment or other property WARNING Calls attention to...

Страница 4: ...ment ART SP2 should be powered from a separate wall outlet with grounding point Attention Avoid use of this machine around Pacemakers It has been shown that electronic appliances including razors hair...

Страница 5: ...y degrade the performance and or may accidentally cause damage to the system Do not dismantle the machine on your own or by uncertified technicians Violation of the requirement may cause harm to the u...

Страница 6: ...or both the patient and the dental surgeon The ART SP2 is a very powerful apparatus There are linear scales of power adjustment for removing prostheses and condensing gutta percha This allows very fin...

Страница 7: ...ents 2 Descriptions of Components 2 Description of Control Panel 4 II Installation Guide 6 III Preparation and Practice before the First Use 9 IV Operation Procedures 11 V Maintenance 13 VI Cleaning a...

Страница 8: ...Handpiece consists of vibrator and a Tip The vibrator is composed of four pieces of piezoelectric plates It generates displacement when driven by the electrical signal from the Main Unit The vibrator...

Страница 9: ...thout interrupting the cleaning process at hand 4 Water Tubing 5 AC Line Socket Seat 6 Micro Motor This is a DC micro motor Micro motor speed will be reacted from input voltage reacted Please see page...

Страница 10: ...ater volume control knob doctors can conveniently adjust water volume based on the operation needs It is noted that not only washing away the dirt left from the calculus and or tartar water here is al...

Страница 11: ...t up It means the operation is under Polishing model During this model the micro motor output power will increase If pushed the button again the operation model will back to scaling model 5 BF symbol...

Страница 12: ...ental limitations apply to both storage and shipping Temperature 0 to 60 Humidity 10 90 at 40 Atmospheric pressure 860 1060 hPa B Labels The meaning of labels printed on the outside of the package box...

Страница 13: ...e unit from the line Remove the cover of the fuseholder by means of a small screwdriver Fit a new fuse of the correct rating and refit the cover of the fuseholder The fuse will be 2A 250V delay action...

Страница 14: ...oper micro motor Several installation suggestions are listed below 1 Because patients may experience some tissue trauma during the treatment it is suggested that the operator use purified or distilled...

Страница 15: ...power intensity by turning the power knob Familiarize yourself with the differences when changing the output power intensity Finally adjust the water volume by turning the water volume control knob F...

Страница 16: ...C Tangential Application E3 TIP Contact on the enamel with the rounded Ttip without pressure when movement 10...

Страница 17: ...hair angle and or position so that you can easily access to the patient s oral cavity 2 Put the saliva ejector into patient s mouth to evacuate away excessive saliva and cleaning remainders 3 Gently m...

Страница 18: ...ained When preparing to execute Polishing model 1 Check the micro motor connection and pushed the MICRO MOTOR change to Polishing model 2 The Forward Reverse that locates rear panel is controlling the...

Страница 19: ...erilized The following suggestions are also useful in extending the product life of the unit A Place the Main Unit in an open area where air can flow freely around it B If need to move the Main Unit h...

Страница 20: ...of the system At the end of the day with the Tip remove the handpiece and cable should be scrubbed with an antiseptic soap or solution and rinsed off with water The handpiece should be scrubbed a seco...

Страница 21: ...reduce the power output Note Replace one new Tip if you find damage worn or weak power of the Tip Main Unit Since the Main Unit does not have direct contact with the patients the cleaning is simple J...

Страница 22: ...AC Power Line x1 3 Three way switch x1 4 PU Tube x1 5 TORQUE WRENCH SET BOX x1 6 Piezo Handpiece x1 7 E5 E2 E3 TIP 8 E type Micro motor Optional Note The above standard outfit is subject to change de...

Страница 23: ...N60555 3 Immunity IEC 1000 4 2 IEC 1000 4 4 IEC 1000 4 5 IEC 801 3 ENV50141 Safety IEC 60601 1 EN1640 2 Function List Scaler power control Water flow control Footswitch control Polisher mode 3 Specifi...

Отзывы: