OPERATION
6
73
SMARTMAC 400
www.krohne.com
07/2022 - 4008875401 - MA SMARTMAC_400 R01 en
6.5 SMARTPAT sensor calibration
To start a sensor calibration proceed as follows:
The further procedure is described sensor-specifically in the following chapters.
6.5.1 SMARTPAT COND sensor calibration
Three calibration methods are available:
"Product calibration" to calibrate the sensor with the help of a reference sensor
"Calib. solution" to calibrate the sensor with the help of a reference solution
"Input cell constant" to enter the calibrated cell constant of a sensor
Product calibration
INFORMATION!
For sensor handling during calibration please consider the standard documentation of the
sensor!
Press > to enter the main menu.
Press
or
until the submenu PAT sensor is highlighted and press > to enter the submenu.
Press > to enter the submenu Calibration.
Press > to enter the submenu Start calibration.
Choose calibration method Product calibration and press
^
.
Choose between Yes and No to set all outputs on hold or not and press
^
.
Press
^
when measured value is stable.
Press
^
to confirm displayed calibration value.
Choose Yes to save value and press
^
to go on with the calibration or choose No to abort calibration.
Choose unit of reference value between µS/cm and mS/cm for measurand (B5.4.1.2) conductivity
orMOhm*cm and kOhm*cm for measurand (B5.4.1.2) resistivity and press
^
.
Set reference value with > (selecting position) and
or
(selecting number). Press
^
for confirmation of
adjusted value.
Press
^
to confirm displayed calibration parameter.
Choose Yes to save parameter and press
^
to go on with the calibration or choose No to abort calibration.
Set calibration date with > (selecting position) and
or
(selecting number). Press
^
for confirmation of
adjusted date.
Choose between Yes and No to set all outputs on hold or not and press
^
.