6
OPERATION
50
SMARTMAC 400
www.krohne.com
07/2022 - 4008875401 - MA SMARTMAC_400 R01 en
B2.1.3.1
Temperature comp.
Setting temperature compensation (only displayed if SMARTPAT PH sensor
is connected)
Options:
•
Automatic
•
Manual
B2.1.3.2
Set temp. comp.
Setting the compensation temperature (only displayed if SMARTPAT PH
sensor is connected and Manual is selected in B2.1.3.1)
000.00…050.00
B2.1.4.1
Buffer solution 1
Measure buffer solution 1 (only displayed if SMARTPAT PH sensor is
connected)
B2.1.4.3
Set buffer solution 1
Setting of buffer solution 1 (only displayed if SMARTPAT PH sensor is
connected)
00.00…14.00
B2.1.5.1
Buffer solution 2
Measure buffer solution 2 (only displayed if SMARTPAT PH sensor is
connected)
B2.1.5.3
Set buffer solution 2
Setting of buffer solution 2 (only displayed if SMARTPAT PH sensor is
connected)
00.00…14.00
B2.1.6.2
Calibration parameter
Calibration parameters are displayed (only displayed if SMARTPAT PH
sensor is connected)
Slope:
Offset:
Message
B2.1.7.1
Save parameter?
Do you want to save parameters? (only displayed if SMARTPAT PH sensor is
connected)
Options:
•
Yes
•
No
B2.1.7.2
Set calibration date
Setting of the calibration date (only displayed if SMARTPAT PH sensor is
connected and Yes is selected in B2.1.7.1)
YYYY-MM-DD
B2.1.8
All outputs on hold?
Primary value is displayed (only displayed if SMARTPAT PH sensor is
connected)
Do you want to set all outputs on hold?
Options:
•
Yes
•
No
B2.2
One point calibration
One point calibration, will be implemented soon (only displayed if SMARTPAT
PH sensor is connected)
B3
Calibration
Calibration of the SMARTPAT ORP sensor (only displayed if SMARTPAT ORP
sensor is connected)
B3.2.2
All outputs on hold?
Do you want to set all outputs on hold? (only displayed if SMARTPAT ORP
sensor is connected)
Options:
•
Yes
•
No
When message "FD (Field Device) in multidrop-mode" is displayed, set
Polling address (B5.5.2) to zero!
Menu level Designation / function
Settings / descriptions