INSTALLATION
3
19
SMARTMAC 400
www.krohne.com
07/2022 - 4008875401 - MA SMARTMAC_400 R01 en
3.6 Wall mounting
3.6.1 Plastic housing
1
Mark the holes with a suitable pen
1
.
2
Drill the 4 holes and fasten the device securely to the wall
2
.
CAUTION!
Always note the following items to ensure a proper and safe installation:
•
Make sure that there is adequate space to the sides.
•
The device must not be heated by radiated heat (e.g. exposure to the sun) to a electronics
housing surface temperature above the maximum permissible ambient temperature. If it is
necessary to prevent damage from heat sources, a heat protection (e.g. sun shade) has to be
installed.
•
Units installed in control cabinets require adequate cooling, e.g. by fan or heat exchanger.
•
Do not expose the unit to intense vibration.
•
Use assembly materials and tools in compliance with the applicable occupational health and
safety directives (assembly materials and tools are not part of the scope of delivery).
CAUTION!
Installation, assembly, start-up and maintenance may only be performed by appropriately
trained personnel. The regional occupational health and safety directives must always be
observed.
INFORMATION!
A mounting system with a minimum load force of 0.1 kN (for example FISCHER type UX10)
suitable for the background has to be applied.
Figure 3-4: Wall mounting of the plastic housing
INFORMATION!
For outdoor mounting, we recommend the optional weather protection cover XGSH010000 for
wall and pipe mounting. For more flexibility and convenient mounting, we recommend wall
mounting brackets XGSH010020 for plastic housing.