6
OPERATION
54
SMARTMAC 400
www.krohne.com
07/2022 - 4008875401 - MA SMARTMAC_400 R01 en
B5.4.1
Measurand
Measurand settings (only displayed if SMARTPAT COND sensor is
connected)
B5.4.1.2
Set measurand
Setting of the measurand (only displayed if SMARTPAT COND sensor is
connected)
Options:
•
Conductivity
•
Resistivity
B5.4.2
Meas. range
Measuring range settings
B5.4.2.2
Lower range value
Set lower range value
Range: 0.0000…14.000 pH (only displayed if SMARTPAT PH sensor is
connected)
Range: 100.000
-3
…20.000 µS/cm (only displayed if SMARTPAT COND sensor
is connected)
Range: -1500.0…+1500.0 mV (only displayed if SMARTPAT ORP sensor is
connected)
B5.4.2.3
Upper range value
Set upper range value
Range: 0.0000…14.000 pH (only displayed if SMARTPAT PH sensor is
connected)
Range: 100.000
-3
…20.000 µS/cm (only displayed if SMARTPAT COND sensor
is connected)
Range: -1500.0…+1500.0 mV (only displayed if SMARTPAT ORP sensor is
connected)
B5.4.3
Time constant
Time constant settings
B5.4.3.2
Set time constant
Set time constant
Range: 01…60 seconds
B5.5
I/O HART
I/O HART settings
B5.5.1
Message
Message setting
B5.5.1.2
Set message
Set message (max. 32 digits)
B5.5.2
Polling address
Polling address setting
B5.5.2.2
Set polling address
Set polling address
Range: 00…63
B5.5.3
Loop curent mode
Loop curent mode setting
B5.5.3.2
Enable current out.?
Do you want to enable the current output?
Options:
•
Yes
•
No
B5.5.4
Device variables
Device variables are displayed (only displayed if SMARTPAT PH sensor is
connected)
B5.5.4.1
pH value
pH value is displayed (only displayed if SMARTPAT PH sensor is connected)
-Min…Max
B5.5.4.2
Max. pH
Max. pH value is displayed (only displayed if PH sensor is connected)
B5.5.4.3
Min. pH
Min. pH value is displayed (only displayed if SMARTPAT PH sensor is
connected)
B5.5.4.4
Electrode voltage
Electrode voltage is displayed (only displayed if SMARTPAT PH sensor is
connected)
-Min…Max
Menu level Designation / function
Settings / descriptions