OPERATION
6
43
SMARTMAC 400
www.krohne.com
07/2022 - 4008875401 - MA SMARTMAC_400 R01 en
A5.3.1
Mode
Operating mode of 2. current output
Options:
•
Off
•
Power Supply
•
Current Output
A5.3.3
Variable selection
Selection of displayed value (only displayed if Current Output is selected in
A5.3.1)
Options:
•
PV
•
SV
•
TV
•
QV
•
4...20 mA (L1)
A5.3.4
Lower range value
Definition of the lower range value (only displayed if Current Output is
selected in A5.3.1)
-Min…Max µS/cm
A5.3.5
Upper range value
Definition of the upper range value (only displayed if Current Output is
selected in A5.3.1)
-Min…Max µS/cm
A5.3.6
Lower ext. range
Definition of the lower ext. range (only displayed if Current Output is selected
in A5.3.1)
Range: 3.8000...20.500 mA
A5.3.7
Upper ext. range
Definition of the upper ext. range (only displayed if Current Output is
selected in A5.3.1)
Range: 3.8000...20.500 mA
A5.3.8
Error function
This function describes, if a lower (3.5 mA) or an upper (21.5 mA) error value
is displayed (only displayed if Current Output is selected in A5.3.1)
Options:
•
Hold value
•
High error current
•
Low error current
A5.3.9
Low error current
Low error current (only displayed if Current Output is selected in A5.3.1 and
Low error current is selected in A5.3.8)
Range: 3.5000...3.6000 mA
A5.3.10
High error current
High error current (only displayed if Current Output is selected in A5.3.1 and
High error current is selected in A5.3.8)
Range: 2.1000...21.500 mA
A5.3.11
Trimming 4 mA
Trimming 4 mA (only displayed if Current Output is selected in A5.3.1)
A5.3.11.2
4 mA Trimming
4 mA Trimming (only displayed if Current Output is selected in A5.3.1)
Range: 0.0000...25.000 mA
A5.3.12
Trimming 20 mA
Trimming 20 mA (only displayed if Current Output is selected in A5.3.1)
A5.3.12.2
20 mA Trimming
20 mA Trimming (only displayed if Current Output is selected in A5.3.1)
Range: 0.0000...25.000 mA
A5.4
Relay 1
Settings for Relay 1
Menu level Designation / function
Settings / descriptions