6
OPERATION
48
SMARTMAC 400
www.krohne.com
07/2022 - 4008875401 - MA SMARTMAC_400 R01 en
B1
Calibration
Calibration of SMARTPAT COND sensor (only displayed if SMARTPAT COND
sensor is connected)
B1.1
Start calibration
Start of SMARTPAT COND sensor calibration (only displayed if SMARTPAT
COND sensor is connected)
B1.1.2.1
Calibration method
Setting of calibration method (only displayed if SMARTPAT COND sensor is
connected)
Options:
•
Product calibration
•
Calibration solution
•
Input cell constant
B1.1.3.2
All outputs on hold?
Setting all outputs on hold? (only displayed if SMARTPAT COND sensor is
connected)
Options:
•
Yes
•
No
When message "FD (Field Device) in multidrop-mode" is displayed, set
Polling address (B5.5.2) to zero!
B1.1.4.1
Value stable?
Is displayed value stable? (only displayed if SMARTPAT COND sensor is
connected and if Product calibration is selected in B1.1.2.1)
Confirm stable value with
^
B1.1.4.3
Calibration value
Calibration value is displayed (only displayed if SMARTPAT COND sensor is
connected and if Product calibration is selected in B1.1.2.1)
B1.1.4.4
Save value?
Do you want to save the calibration value? (only displayed if SMARTPAT
COND sensor is connected and if Product calibration is selected in B1.1.2.1)
Options:
•
Yes
•
No
B1.1.4.5
All outputs on hold?
Primary value is displayed (only displayed if SMARTPAT COND sensor is
connected, if Product calibration is selected in B1.1.2.1 and No is selected in
B1.1.4.4)
Do you want to set all outputs on hold?
Options:
•
Yes
•
No
B1.1.4.6
Set unit of ref. val.
Set unit of reference value (only displayed if SMARTPAT COND sensor is
connected, if Product calibration is selected in B1.1.2.1 and Yes is selected in
B1.1.4.4)
Options:
•
mS/cm (only displayed if Conductivity is selected in B5.4.1.2)
•
µS/cm (only displayed if Conductivity is selected in B5.4.1.2)
•
kOhm*cm (only displayed if Resistivity is selected in B5.4.1.2)
•
MOhm*cm (only displayed if Resistivity is selected in B5.4.1.2)
B1.1.4.7
Set reference value
Setting of the reference value (only displayed if SMARTPAT COND sensor is
connected, if Product calibration is selected in B1.1.2.1 and Yes is selected in
B1.1.4.4)
Range: 0.0000…Max.
Menu level Designation / function
Settings / descriptions