Man242 (1 Sep. 2015)

WHM-C10D

Digital Feedback Temp. Controller

Certifications by 

- - -

CE Marked Products.

ISO 9001 Certified.
PL Insurance.

Patented(No. 10-0554098)  Jog Shuttle Control System.

Operation Manual 

(Version 2.1.1)

written for

Model WHM-C10D

WHM-C10D Models

DAIHAN-Sci.com or 

ALL

for

LAB.com Shopping Mall

Sales & Basic A/S :

By the Nearest Official Distributors.

²

English Page : P. 1 ~ 11

²

Korean Page : P. 12 ~ 27

Содержание WHM-C10D

Страница 1: ...ISO 9001 Certified PL Insurance Patented No 10 0554098 Jog Shuttle Control System Operation Manual Version 2 1 1 written for Model WHM C10D WHM C10D Models DAIHAN Sci com or ALL for LAB com Shopping M...

Страница 2: ...ler WHM C10D to which this declaration relates is in conformity with the following standard or other normative document EN 61326 EN 61000 3 2 EN 61000 3 3 EN 61010 1 EN 61010 2 following the provision...

Страница 3: ...IME Jog Shuttle switch Turn Push l User s self compensation function to control difference between Real Temp and l Indicated Temp 10 0 l Digital LCD with Back Light Function l Storage function Timer A...

Страница 4: ...Mall DAIHAN Scientific WHM C10D Digital Feedback Temp Controller 4 27 Appearance 3 The appearance of Digital Feedback Controller WMH C10D is shown in Figure 1 Figure 1 Appearance of WMH C10D Figure 2...

Страница 5: ...Multifunctional Digital Feedback Controller Figure 2 Controller of Digital Feedback Controller Operation Status on LCD Display Indication of Set Value Indication of Current Value LED indicating Heatin...

Страница 6: ...sage of the external sensor SS300 optional K type sensor or SS310 optional Teflon coated K type sensor A If you don t connect K type sensor the temperature is controlled by the measured value of Heati...

Страница 7: ...shuttle for adjustment of temperature the set value is change firstly and the current value goes after the set value Then the current value is reached to the set value and it means that Digital Feedba...

Страница 8: ...nd sample with the measured value by the User A Press Mode Switch and Jog Shuttle simultaneously in the Heating Mode Mode S W Jog S W B In this status it is changed into Offset Mode and the user can a...

Страница 9: ...ating and touching this surface may cause a serious injury If you need to touch it you should have sufficient cooling time down to 70 on the LCD display before touching The recommended power supply is...

Страница 10: ...ct is designated for separate collection at an appropriate collection point Do not dispose of as household waste This product is required to comply with the European Union s Waste Electrical Electroni...

Страница 11: ...unctions before shipment 2 This instrument is warranted for two years from purchasing date 3 A breakdown caused by customer s misuse or natural disaster is excepted from this warranty 4 This certifica...

Отзывы: