Installation and operating instructions
IFC 020
5/6
→
VALUE P = set pulse value per unit volume
(appears only when “PULSE/VOL.” set under “SELECT P”, press
→
key.
• XXXX PulS/m3
• XXXX PulS/Liter
• XXXX PulS/US.Gal
• XXXX PulS/
user unit, factory-set:
“Liter”
or
“US MGal”,
see Sect. 5.12.
Select using
↑
or
↓
key.
Transfer to number setting with
→
key. 1st digit (cursor) flashes.
Set numerical value
• XXXX
(setting range depends on pulse width and
full-scale range:
P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
)
Change flashing digit (cursor) with
↑
or
↓
key,
shift cursor 1 place to right with
→
key.
Press
↵
key to return to Fct. 1.06 PULS. P.
or
Please refer to Sect. 3.2 “factory settings”
→
VALUE P = set pulse value per unit time,
(appears only when ”PULSE/TIME” has been set under “SELECT P”),
press
→
key.
• XXXX PulSe/Sec
• XXXX PulSe/min
• XXXX PulSe/hr
• XXXX PulSe/
user unit, factory-set:
“hr”
, see Sect. 5.12.
Select using
↑
or
↓
key.
Transfer to number setting with
→
key, 1st digit (cursor) flashes.
Set numerical value
• XXXX
(setting range depends on pulse width)
Change flashing digit (cursor) with
↑
or
↓
key,
shift cursor 1 place to right with
→
key.
Press
↵
key to return to Fct. 1.06 PULS. P.
Refer to Sect. 2.4 for connection diagrams, and to Sect. 5.14 for characteristics.