39
Notes
• Prior to the start of measurement, be sure to perform white calibration. For details, refer to “White
Calibration and Gloss Calibration” on P.36.
• To display the color difference, color difference target must be set before measurement.
• To measure a target, select the target number before measurement.
• For accurate measurements, measure under the same conditions (surrounding temperature, etc.).
Operating Procedure
1
Press [
] to bring up the [Target] or [Sample]
screen, depending on the objective.
The <Target> screen or the <Sample> screen
is displayed.
Note: If the menu screen is displayed, press
[ESC]
and perform
the operation after the Results Display screen is displayed.
2
Set the specimen measuring port on the sample to
measure.
• Take care to prevent floating or tilting from occurring.
• Using the optional stapler type target mask (CM-A268) makes it possible to determine the measurement
position more accurately.
Measuring
MEASURING
To measure concave surfaces, as
shown in the following figure, use the
standard accessory flat type battery
cover (CM-A218).
3
Press the measurement button.
The specimen is measured and the results are displayed on
the screen.
• The selected number will be used as the target data
number. If data is already present for that number, a
confirmation dialog will ask if the data should be
overwritten. Press the
[Confirmation] key
to overwrite.
• The sample No. is automatically assigned sequentially in
the order of measurements.
Memo
• When the number of sample data sets stored in the memory reaches 5,100, “The device memory is full. Please delete
some data to measure.” will be displayed, indicating that some data sets must be deleted before measurement is
possible.
Measurement
Содержание CM-23d
Страница 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Страница 30: ...27...
Страница 33: ...30 Measurement 2 Attach the Wrist Strap Attaching the Wrist Strap Wrist strap attaching position Preparation...
Страница 82: ...79...
Страница 111: ...108 Setting 3 MEMO...
Страница 112: ...109...
Страница 132: ...129...
Страница 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Страница 138: ...135...
Страница 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Страница 142: ...139 Dimensions Unit mm...
Страница 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...