134
133
Troubleshooting
5
Symptom
Check Point
Action
Data cannot be
transferred to a computer
from the instrument.
No commands from the
computer are being
accepted.
Is the USB cable connected
correctly?
Properly connect the instrument’s USB connecting
terminal to the computer’s USB port with the USB
cable supplied with the instrument.
Has Bluetooth communication
been established?
Install and connect the Bluetooth module correctly.
Printing is not possible.
Has Bluetooth communication
been established?
Install and connect the Bluetooth module correctly.
Samples and settings are
not held in memory and
disappear immediately.
The instrument’s backup
battery may be low
immediately after purchase or
following a period of
prolonged non-use. Turn the
instrument ON to charge the
backup battery. Under this
condition, the backup battery
will be fully charged in
20 hours.
The backup battery in the instrument should last
10 years under normal use. If the instrument fails to
retain data in memory even after the battery has been
fully charged, however, the battery has likely reached
the end of its life and requires replacement.
To replace the backup battery, please contact a
KONICA MINOLTA-authorized service facility.
Содержание CM-23d
Страница 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Страница 30: ...27...
Страница 33: ...30 Measurement 2 Attach the Wrist Strap Attaching the Wrist Strap Wrist strap attaching position Preparation...
Страница 82: ...79...
Страница 111: ...108 Setting 3 MEMO...
Страница 112: ...109...
Страница 132: ...129...
Страница 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Страница 138: ...135...
Страница 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Страница 142: ...139 Dimensions Unit mm...
Страница 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...