35
2
Use [
] or [
] to move the cursor to “Calibration
(Including Zero Cal.)”, and then press the
[Confirmation] key.
3
Set the instrument in the calibration stage to
measure the calibration stage’s zero calibration
hole.
4
Press the measurement button.
Zero calibration will be performed.
Notes
• Do not move the instrument until zero calibration is
complete.
When zero calibration is completed, a screen prompting
white calibration is displayed. Continue to step 3 on the
following page to perform white calibration.
Please set the
instrument on the zero
calibration box and
press the measurement
button.
Calibration stage
Place the instrument
on the calibration stage
according to the marks.
Calibration stage
Place the instrument
on the calibration stage
according to the marks.
Calibration stage
Place the instrument
on the calibration stage
according to the marks.
Calibrating to zero
CALIBRATION
Содержание CM-23d
Страница 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Страница 30: ...27...
Страница 33: ...30 Measurement 2 Attach the Wrist Strap Attaching the Wrist Strap Wrist strap attaching position Preparation...
Страница 82: ...79...
Страница 111: ...108 Setting 3 MEMO...
Страница 112: ...109...
Страница 132: ...129...
Страница 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Страница 138: ...135...
Страница 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Страница 142: ...139 Dimensions Unit mm...
Страница 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...