68
67
Measur
ement
2
Tolerance Settings
Specify the tolerance used for pass/fail judgment of measured for each target.
Operating Procedure
Start the procedure from the <Target menu> - <Pass/Fail> screen.
1
Use [
▲
] or [
▼
] to move the cursor to “Edit
tolerance”, and then press the [Confirmation] key.
The tolerance list screen is displayed.
2
Use [
▲
] or [
▼
] to move the cursor to the item to set,
and then press the [Confirmation] key.
The <Edit tolerance> screen
is displayed.
3
Use [
▲
]/[
▼
] and [
◀
]/[
▶
] to move the cursor to the
item to set, and then press the [Confirmation] key to
change the setting.
• When the selected item is not currently checked,
pressing the
[Confirmation] key
causes the item to be
checked, after which the setting value can be changed.
•
Press [
◀
] or [
▶
]
to move the cursor to the setting value
area. Pressing the
[Confirmation] key
causes the
cursor to appear on the value. Press
[
▲
] or [
▼
]
to
change the value.
Move the cursor between the digits of the value by
pressing
[
◀
] or [
▶
]
.
Settings
{
x / y
: -0.2000 to 0.2000
{
Equation / MI
: 0.00 to 20.00
{
Other than the above : -20.00 to 20.00
• Press the
[Confirmation] key
to confirm every time the
setting of an item is changed.
4
After all settings have been configured, use [
▲
] or
[
▼
] to move the cursor to “OK”, and then press the
[Confirmation] key.
The setting is confirmed and the screen returns to the
<Pass/Fail> screen
.
Pressing the
[ESC]
while configuring the settings will prevent
the settings on the currently opened screen from being
applied and revert all settings to the previous condition.
5
Press [ESC] to return to the <Pass/Fail> screen.
Содержание CM-23d
Страница 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Страница 30: ...27...
Страница 33: ...30 Measurement 2 Attach the Wrist Strap Attaching the Wrist Strap Wrist strap attaching position Preparation...
Страница 82: ...79...
Страница 111: ...108 Setting 3 MEMO...
Страница 112: ...109...
Страница 132: ...129...
Страница 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Страница 138: ...135...
Страница 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Страница 142: ...139 Dimensions Unit mm...
Страница 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...