36
35
Measur
ement
2
White Calibration
A message prompting white calibration is displayed on the instrument after the power is turned ON.
Memo
• If the calibration interval is turned ON and a time is configured, a message prompting white calibration will be
displayed the next time the power is turned ON or when measurement is performed after the set time has passed
since the previous white calibration. (Refer to P.124 “Calibration Interval Messages”)
• The reading may fluctuate slightly due to changes in the ambient temperature or due to heat generation caused by
repeated operation of the instrument. In such cases, perform white calibration regularly.
Notes
• White calibration must be performed at the same temperature at which measurement will be performed.
• Perform white calibration after the instrument has had time to adapt to the ambient temperature.
Operating Procedure
Start the procedure from the measurement screen.
Although white calibration can be performed from the prompt when the power is turned ON and from the screen
following zero calibration, the following explains the procedure to perform white calibration from the
measurement screen.
1
Press [MENU], and then use [
◀
] or [
▶
] to display the
<Calibration> screen.
Notes
• The following screen prompting calibration is
displayed when the instrument starts up. If zero
calibration has not been performed, the cursor will
appear on “Calibration (Including Zero Cal.)”.
Otherwise, the cursor will appear on “Calibration
(Excluding Zero Cal.)”.
2
Use [
▲
] or [
▼
] to move the cursor to “Calibration
(Excluding Zero Cal.)”, and then press the
[Confirmation] key.
Calibration
(Excluding Zero Cal.)
Calibration
(Including Zero Cal.)
Calibration
Zero calibration date
2019/11/19 08:44:31
White calibration date
2019/11/19 09:41:33
Содержание CM-23d
Страница 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Страница 30: ...27...
Страница 33: ...30 Measurement 2 Attach the Wrist Strap Attaching the Wrist Strap Wrist strap attaching position Preparation...
Страница 82: ...79...
Страница 111: ...108 Setting 3 MEMO...
Страница 112: ...109...
Страница 132: ...129...
Страница 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Страница 138: ...135...
Страница 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Страница 142: ...139 Dimensions Unit mm...
Страница 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...