Make sure that the measurement ball valve is shut off.
•
Remove transport protection cap of the head of the sensing probe.
•
Mount the sensing probe in the measurement valve with shut-off function in such a way that the
alignment pin fits in the recess on the measurement ball valve.
•
Screw the retainer nut by hand so far that a certain resistance is noticeable.
•
Check the correct installation position of the flow meter. The alignment pin must fit in the recess on
the measurement ball valve.
•
Tighten the red coloured retainer nut by hand. Tightening by hand should be sufficient. However, if
the seal is not leak tight carefully tighten the nut with an appropriate tool a bit further.
•
The mechanical installation of the flowmeter is therewith completed.
Flow Direction
3.5.4. Installation of the flow meter sensing probe
The flow direction is indicated with an arrow on the tip of the probe. Due to the alignment pin is the installation
of the sensing probe in the measurement ball valve only possible in the direction of the flow. After a removal,
the sensing probe will be re-installed in the measurement section in exactly the same position as done at the
factory. Hence, the highest reproducibility is guaranteed.
3.5.5. Removal of the sensing probe of the flowmeter
•
Shut off the measurement ball valve (see page 11).
•
Turn off the power supply, remove the cover and disconnect the power wires on the screw terminal.
•
Loose the retainer nut and pull the sensor probe from the measurement section.
•
Operation without the flowmeter installed page 11.
Never remove the flow meter while the measurement ball valve is open.
That is extremely dangerous!