![Kobold Heinrichs Coriolis Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/kobold/heinrichs-coriolis/heinrichs-coriolis_operating-manual_1982428094.webp)
Page 94 of 105
OPERATING MANUAL HPC & UMC4
15. Density calibration
For continuous processes, which process only small variations in temperature and liquid media of compa-
rable density, a density calibration can be carried out locally.
15.1 Conditions
For a local density calibration the following conditions must be fulfilled:
The sensor must be suitable for a density measurement. For all these sensors the manufacturer
offers a 3-point calibration with an accuracy of at least 5 g/l. Sensors, for which the manufacturer
calibration is not offered, are also not suitable for local single point calibration.
For the local calibration the medium’s density must be known or be determined by a suitable pro-
cedure using exactly the same process temperature present in the sensor during measurement.
The medium must be liquid. The one-point density calibration is not suitable for gaseous media.
15.2 Procedure
The procedure of the density calibration is as follows:
First the transmitter is switched on and the sensor is filled with the medium under operating con-
ditions.
So that the sensor including housings and flanges can take the operating temperature a waiting
period of at least 15 minutes is to be kept.
After the input of the customer’s password select functional class DENSITY and “density calibra-
”. (See also chapter 14.6.15 on page 68.)
Independently of the kind of the liquid medium and its temperature the beginning of the calibra-
tion is confirmed by selecting “yes”. Thereupon the transmitter measures medium temperature
and current resonant frequency of the sensor.
In the following the function “measured values hot medium“ has to be selected The displayed
measured values of temperature and frequency are confirmed by 2 times pressing the Enter-key.
In the next field the density of the medium has to be entered in unit of g/l or kg/m³. (See also
chapter 14.6.16 on page 68.)
Subsequently, the function “finish density calibration” is used to finish density calibration. (See al-
so chapter 14.6.17 on page 68.)
Finally the function “density measurement on/off” activates the density measurement. (See also
Now the measured density and also volume flows can be indicated or assigned to one of the out-
puts e.g. current output 2.