OPERATING MANUAL HPC & UMC4
12.2 Weight
Approx. 2.4 kg (5.5 lbs) (separate UMC4 transmitter without mounting system)
12.3 Material
Enclosure: Painted aluminum pressure die
–casting, max.0.5% Mg; yellow chromating.
12.4 End connection
To connect the remote sensor to the transmitter, a special connection cable must be used.
For further details see Sections 4.6
“Connection to the transmitter” on page 21, and 12.5.1.2 Connec-
tion of the sensor on page 37.
12.5 Electrical connections and their protection classes
Auxiliary power
90 V - 265 V AC
50/60 Hz
24 V AC
+ 5 %,
− 20 %
50/60 Hz
19 V to 36 V DC
Power input
4.5 VA
Main fuse:
5x20 mm IEC 60127-2
Main voltage
rated Current
rated voltage
breaking capacity
90V ... 265V AC
250mAT
250V AC
1500A / 250V AC
24V AV
250mAT
250V AC
1500A / 250V AC
19V ... 36V DC
250mAT
250V AC
1500A / 250V AC
Supplier
: Little Fuse Series 0215.250 HXP
12.5.1 Wiring diagrams
12.5.1.1 Power connections
Designation
Terminal / Pin designation
Type of protection
Standard
Ex ia
(Non Ex)
Power supply
L
or “+”
L
x
N
or “-“
N
x
PE
PE
x
Power connection terminals