OPERATING MANUAL HPC & UMC4
Page 71 of 105
14.7.4 Temperature limit MIN
The MIN limiting value for temperature can be evaluated via the status output. This value is entered in the
set temperature unit.
After selecting the
Temperature limit MIN
function, press
„Enter“ to display the following selection field:
MIN temperature
-010 °C
The current MIN limiting value will be displayed. If the measured value falls below the limiting value, the
“Alarm” status message will be displayed. As mentioned in Section 13.4.3.2 Input window/modify a value,
the operator can change the MIN limiting value for temperature measurement.
14.7.5 Temperature limit MAX
The MAX limiting value for temperature can be evaluated via the status output. This value is entered in
the set temperature unit.
After selecting the
Temperature limit MAX
function, press
„Enter“ to display the following selection field
MAX temperature
+ 250 °C
The current MAX limiting value will be displayed. If the measured value falls below the limiting value, the
“Alarm” status message will be displayed. As mentioned in Section 13.4.3.2 Input window/modify a value,
the operator can change the MAX limiting value for temperature measurement.
14.7.6 Max. measured temperature
After selecting this display, the largest measured temperature will be displayed. For comparison, the set
maximum limiting value will be displayed in the first line.
permissible +250 °C
measured +197 °C
This value cannot be reset since it stores the maximum measured process temperature.