OPERATING MANUAL HPC & UMC4
14.3.1 Unit of counters
After choosing the
Unit of counters
function and pressing
„Enter“, the current forward and reverse counter
unit will be displayed:
Accumulation of:
[kg]
According to the description in Section 13.4.3.1 Selection window/make a selection, one of the following
units can be selected.
Mass units:
g, kg and t as well as lbs, shton, lton and oz
or
Volume units: m³, cm³ and l as well as USG, UKG, USB, ft³, acf and floz
Programmable mass unit:
xxQM,
Programmable volume unit:
xxQV.
When the unit is changed, the counters will be reset to 0.00 automatically.
The volume unit only makes sense if the sensor has been calibrated for density measurement. Press
„Enter“ to confirm and save the selection. Forward and reverse counters will now show the selected unit.
The valence of the programmable units are defined by the settings of the flow units described in sections
14.5.2
“Factor mass flow QM programmable unit” on page 59 and 14.5.8 “Factor volume flow QV pro-
14.3.1 Reset counters
To reset the totalizing counters, you definitely need to toggle to [yes]. Forward and reverse counters will
be reset at the same time (0.00).
Reset counters
[no]
According to the description in section 13.4.3.1 Selection window/make a selection, one of the indicated
units can be selected. By pressing Esc or toggling to [no] the operator can cancel the current action with-
out changing the counter readings.