Application Examples
35
AZURA V 2.1S User Manual, V6860, Version 2.2
Backflushing the Column
The backflushing of a column is used, for example, in the following cases:
analysis of heavily retarding substances
optimization of analysis times
flushing of a column
Note:
The eluent flushes the column after switchover in opposite direction.
Make sure there is no pressure when switching so that the column is not dam-
aged.
Backflushing the Precolumn
The precolumn is used in chromatography, for example, in the following cases:
preseparation or separation of heavily retarding substances
Functional Principle
From the valve, the column
1
is filled with sample through the channel
2->3
.
The quicker part of the substances is separated through the channels
6->1
and
5->4
and transported to the detector
2
.
Fig. 24 backflushing - transport of the sample onto the column
After switching over the valve, heavily retarding substances are eluted in oppo-
site direction from the column and are transported back to the detector
2
.
Fig. 25 backflushing of the column
1
2
3
4
5
6
1
2
1
2
3
4
5
6
2
1