Installation
AZURA V 2.1S User Manual, V6860, Version 2.2
BIN 0 - 3
If the DIP switches were set to binary operation, then the connections BIN 0 -
BIN 2 are available as inputs. The output BIN 3 is active for the ports 1-6.
On the 12 port and 16 port valves, BIN 3 is defined as an input so that the ports
9-12 or 9-16 can be used.
Note
The binary input is inactive during the change of the valve port. This can be
used to control further external devices.
Binary Code
A binary code is entered during binary control so that the valve can be set exter-
nally in the correct position (nominal position).
Prerequisites
The DIP switch of the device was set for the binary control.
.
Signal
Explanation
GROUND
Ground connection for start and error signals
HOME/POS. 1 Connection for short circuit for creating a switch signal:
Valve drive is set to position 1.
FORWARD/
LOADING
Connection for short circuit for creating a switch signal, exam-
ple for output:
LOADING
Move to the next higher port of the valve, e. g. from posi-
tion 2 to position 3
BACKWARD/
INJECTING
Connection for short circuit for creating a switch signal, exam-
ple for output:
INJECTING
Move to the next lower port of the valve, e. g. from position
6 to position 5
OUT
Connection for short circuit (no current)
INJECTING, port 1, low-impedance
LOADING, port 2-6 or 2-16, high-impedance
Position
BIN 0
(2
0
=1)
BIN 1
(2
1
=2)
BIN 2
(2
2
=4)
BIN 3
(2
3
=8)
Result,
binary
1
0
0
0
0
0
2
1
0
0
0
1
3
0
1
0
0
2
4
1
1
0
0
3