Safety
13
AZURA V 2.1S User Manual, V6860, Version 2.2
Participation during an installation of a device or a training by the company
KNAUER or an authorized company.
If you do not belong to this or a comparable professional group, you may not
perform the work described in this user manual under any circumstances. In this
case, please contact your superior.
Safety Equipment
When working with the device, take measures according to lab regulations and
wear protective clothing:
Safety glasses with side protection
Protective gloves
Lab coat
What must be taken into account?
All safety instructions in the user manual
The environmental, installation, and connection specifications in the user
manual
National and international regulations pertaining to laboratory work
Original spare parts, tools, and solvents made or recommended by KNAUER
Good Laboratory Practice (GLP)
Accident prevention regulations published by the accident insurance compa-
nies for laboratory work
Filtration of substances under analysis
Use of inline filters
Once they have been used, never re-use capillaries in other areas of the HPLC
system.
Only use a given PEEK fitting for one specific port and never re-use it for
other ports. Always install new PEEK fittings on each separate port.
Follow KNAUER or manufacturer's instructions on caring for the columns
More safety-relevant information is listed below:
flammability: Organic solvents are highly flammable. Since capillaries can
detach from their screw fittings and allow solvent to escape, it is prohibited
to have any open flames near the analytical system.
solvent tray: Risk of electrical shock or short circuit if liquids get into the
device's interior. For this reason, place all bottles in a solvent tray.
solvent lines: Install capillaries and tubing in such a way that liquids cannot
get into the interior in case of a leak.
leaks: Regularly check if any system components are leaking.
power cable: Defective power cables are not to be used to connect the
device and the power supply system.