Installation
17
AZURA V 2.1S User Manual, V6860, Version 2.2
Protect the device against direct exposure to sunlight.
Set up the device at a location not exposed to air drafts (A/C systems).
Do not set up the device in the vicinity of other machines that cause floor
vibrations.
Lateral distance to other devices:
At least 5 cm if another device is set up on one side
At least 10 cm if further devices are set up on both sides
Leave the power plug on the rear of the device accessible to be able to
disconnect the device from the mains.
Installation Site
air humidity: below 90 % (non-condensing)
temperature range: 4–40 °C; 39.2–104 °F
Setup
Prior to setting up the device, find a suitable place according to the require-
ments and remove the packaging.. The setup requirements and a description
can be found in the following section.
Unpacking
Prerequisites
Check packaging for damage caused during transportation.
Tools
utility knife
Bruises
Damage to the device by carrying or lifting it on protruding housing parts. The
device may fall and thus cause injuries.
Lift the device on the side of the housing only.
Process
1. Set-up the package in such a way that you can read the label.
2. Using the utility knife, cut the adhesive tape and open the packaging.