22
Installation
AZURA V 2.1S User Manual, V6860, Version 2.2
Description for Stainless Steel Sample Loop
Prerequisites
The ends of the sample loop were cut straight at a right angle to the capillary
axis.
The screw fittings are at hand.
4. Fixate the first seal by screwing it
into port 2. Afterwards unscrew
again.
5. Fixate the second seal in port 5 and
afterwards unscrew again.
6. Screw the sample loop evenly and
simultaneously into port 2 and port
5.
Fig. 10 scheme of an installed sample loop
Process
Figure
1. Insert one end of the sample loop
into the screw fitting
1
.
2. Push the ferrule onto the capillary
so that the broader sight is oppo-
site the screw fitting.
3. Repeat the first steps on the other
end of the capillary.
Fig. 11 screw fitting and ferrule
4. Fixate the first seal by screwing it
into port 2. Afterwards unscrew
again.
5. Fixate the second seal in port 5 and
afterwards unscrew again.
6. Screw the sample loop evenly and
simultaneously into port 2 and port
5.
Fig. 12 scheme of an installed sample loop
Process
Figure
1
2
3
4
5
6
1
2
1
2
3
4
5
6
Содержание Azura V 2.1S