22 Operation
AZURA® Conductivity Monitor CM 2.1S Instructions, V6710
5.2.1 Operating elements
Display and keypad
Explanation
Fig. 11:
Status display
The status display shows the
conduc tivity (line 1), the current
gradient composition (line 2, left),
and the pH value (line 2, right).
Fig. 12:
Arrow keys
Press both arrow keys
simultane ously to activate the
keypad.
Hold down 1 arrow key to
scroll with the other arrow key
through the menus.
Press any one of the arrow keys
to set values and to change
settings.
5.2.2 Menu items and factory settings
Display content Description
The measuring signal can be adjusted with an offset
within a set range between -10 % and +10 %.
Factory setting:
0.0 %
Fig. 13:
By entering a non-zero value at
TEMP-COR, the tem-
perature compensation function of the instrument
is activated. This allows the specific conductivity to
be automatically calculated and output for a tempe-
rature set under
TEMP-REF, regardless of the actual
temperature of the medium flowing through the CM
flow cell.
Factory settings:
TEMP-REF: 25°C
TEMP-COR: 0.0 %
Fig. 14:
Use the menus
Factor 1-4 to define up to 4 calibrat-
ion points. Enter the conductivity or the correction
factor for each point to generate the correct cali-
bration function.
Factory settings:
Factor 1: 12 mS/cm, 100%
Factor 2: 111mS/cm, 100%
Factor 3: 211 mS/cm, 100 %
Factor 4: 299 mS/cm, 100%