Operation 23
AZURA® Conductivity Monitor CM 2.1S Instructions, V6710
Display content Description
Two menus help to calibrate the pH monitor:
#0 is preset to pH07
#1 can be adjusted to a value between pH02....
pH12 and corresponds to the pH value of the
second calibra tion solution that is used.
Select this menu to view the serial number of the
monitor.
Select this menu to view the serial number of the
flow cell.
Select this menu to view the operating hours of the
moni tor.
Use this menu to select an interface for the PC
connection:
LAN DHCP
9600 Baud (RS-232)
Use this menu to view the progress of the program-
med salt or buffer gradient in the status display. To
do so, it is neces sary to enter the conductivity of the
initial concentration of the programmed gradient
first.
In this menu, enter the conductivity of the final
concentra tion of the programmed salt or buffer
gradient.
5.3 Selecting the interface
To view the measuring results of the conductivity monitor via a software,
the interface you use must be selected on the device:
PC: LAN DHCP
PC: 9600 Baud (RS-232)
A physical connection has been established.
The monitor has been switched on.
Prerequisites