12 Installation
AZURA® Conductivity Monitor CM 2.1S Instructions, V6710
4.2 CM flow cell
The CM flow cell is supplied with built-in capillaries that cannot be disas-
sembled. In the event that the capillary is clogged and cannot be cleaned
even by repeated flushing processes, the CM flow cell must be replaced.
There is a plug on the rear side of the CM flow cell, which is used to con-
nect the flow cell to the front side of the monitor.
Note:
The analytical flow cell A4156 generates a certain back pressure
(approximately 1 bar at 1 ml/min). For minimal backpressure choose flow
cell A4157.
Legend:
1
Fitting
2
SST ferrule
3
PEEK ferrule
4
1/8"-1/16"-
reducer
5
1/16"-1/16"-coupling
6
Capillary
7
Indication of manufactu-
rer, part number, serial
number
8
Indication of recom-
mended operating
range
9
Drill holes for mounting
bl
Screws
bm
Plug
Fig. 4:
CM flow cell: front with labeling, accessories
Fig. 5:
CM flow cell: rear side with connector
7
4
bl
1
2
9
6
3
5
8
bm