20 Installation
AZURA® Conductivity Monitor CM 2.1S Instructions, V6710
4.5.2 Firmware Wizard: Setting a dynamic IP address
The device is switched on.
Firmware Wizard is installed and running.
Process
1.
In Firmware Wizard, cleck <Reset LAN Settings...>.
2.
The window <Device connection settings> opens. Enter MAC
adress of the device into the text field <Target device serial num-
ber>
1
.
3.
Select option <Obtain an IP address automatically>
2
.
4.
Click <Reset Conn. Settings>
5
to accept changes.
5.
Restart the device (recommended).
The device is now accessible via a dynamic IP address.
4.6 Power connection
The devices are equipped with an universal switched-mode power supply
rated for 100–240 V AC. The switched-mode power supply automatically
selects the correct supply voltage.
Supply voltage: 115 ± 15 % or 230 ± 10 % VACs
Frequency: 50/60 Hz
Legend:
1
On /Off switch
2
LED
Fig. 10:
Mains power connection
1
2
Electric shock
Electric shock hazard caused by an improperly grounded power
connection
.
Î
Ground the power connection according to the pertinent regulations.
Î
Use a three-conductor line cord.
Prerequisites
Procedure
Result