ASM 2.1L user manual V6815
Initial startup
16
Integrator output
The integrator output sends measuring signals from the detector.
non-bipolar
1 channel
0–5 V
DAC 16 bit
Scalable
Adjustable to offset
Initial startup
Optional mounting brackets, a holder for the Mobile Control or AZURA®
Click can be attached to ASM 2.1L during initial startup (see “Repeat
orders” on page 28). Please check the user manuals of the micro devices
to find all information on initital startup (see “Device overview” on page 5).
PEEK fittings
Component defect
Damage to the ports caused by strongly tightened fittings.
Use 5 Nm torque for stainless steel fittings.
Use 0.5 Nm torque for PEEK fittings.
ASM 2.1L as Compact HPLC isocratic
Tools
Tool kit A1033
+24 V
Event-controlled switching of 24 V
against GND
Max. current: 200 mA
GND
Reference point of the voltage at the signal inputs
Connection
Function
Fig. 6
Scheme of an isocratic compact HPLC system
Procedure
Process
1. Connect the solvent bottles with degasser.
2. Connect the degasser with pump.
3. Connect the pump to the manual valve.
4. Connect the valve with column.
5. Connect the column to the detector.
6. Connect the detector to the waste bottle.