ASM 2.1L user manual V6815
15
Unpacking and setup
Connection
Function
EV 3 (Event 3)
Relay contact
The contact is on a floating basis. Its setting
depends on the settings in the software.
Steady-rate signal:
passive = open relay contact
active = closed relay contact
Pulse:
Closed relay contact for at least 1000 ms
Permissible load of the relay contact: 1 A/ 24 V DC
EV 2 (Event 2)
OC output limited to 5 V
Levels:
passive 5 V
active 0 V
Pulse:
0 V for at least 1000 ms
EV 1 (Event 1)
OC output limited to 5 V
Levels:
passive 5 V
active 0 V
Pulse:
0 V for at least 1000 ms
START IN
TTL input
Low active
Secure switching threshold at least 1000 ms
After receiving a signal (short-circuit to ground) from
an external device, the device starts. If controlled with
software, an electronic trigger is send through the
LAN.
ERROR IN
TTL input
Low active
Secure switching threshold at least 10 mA
After receiving a signal (short-circuit to ground) from
an external device, an error message appears and the
device stops.
ERROR OUT
OC output limited to 5 V
Levels:
passive 5 V
active 0 V
Pulse:
0 V for at least 1000 ms
+5 V
Provides a voltage of 5 V with respect to GND. This
makes it possible to supply a consumer that is
switched by an EVENT.
Max. current: 50 mA
GND
Reference point of the voltage at the signal inputs