ASM 2.1L user manual V6815
13
Unpacking and setup
Unpacking and setup
Contacting the Technical Support
You have various options to contact the technical support:
Phone
+49 30 809727-111
Fax
+49 30 8015010
[email protected]
You can make your requests in English and German.
Location requirements
Only if the requirements for ambient conditions of the operating environ-
ment are met, can the intended use be ensured. Details on the operating
conditions can be found in the Technical Data section.
Device defect
The device overheats at exposure to sunlight and insufficient air circula-
tion. Device failures are very likely.
Set up the device in such a way that it is protected against exposure to
direct sunlight.
Keep at least 15 cm clear at the rear and 5–10 cm at each side for air cir-
culation.
Space Requirements
At least 5 cm if another device is set up on one side
At least 10 cm if further devices are set up on both sides
At least 15 cm on the rear panel for the fan.
Leave the power plug on the rear of the device accessible to be able to
disconnect the device from the mains.
General requirements
Position the device on a level and even surface.
Protect the device against direct exposure to sunlight.
Set up the device at a location not exposed to air drafts (A/C systems).
Do not set up the device in the vicinity of other machines that cause
floor vibrations.
Power supply
The device is only suitable for alternating current. The supplied power
cable is to be used to connect the device to the mains supply.
Unpacking
Note:
The leak sensor may malfunction if the device stands on a tilted sur-
face. Use a level to check that the device stands horizontally.
Store all packing materials. Included packing list should be kept for repeat
orders.
Damage to the leak tray or front cover possible while
carrying, setting up and installing a device. Grip the
device at its sides near the middle when lifting or mov-
ing.
Electrostatic-discharge hazard
Symbol
Meaning