Device overview
Configuration of modules
Maximum number of same modules
Three modules can be integrated in an assistant. The number of same
device types is limited:
1x Degasser
1x Detector UVD 2.1S
2x Pumps (only with ClarityChrom
®
or PurityChrom
®
), HPG is not sup-
ported
Limitation of combination
The pump P 4.1S cannot be placed to the right of detector UVD 2.1S,
because the fittings of the detector flow cell cannot be mounted.
An assistant with only degasser integrated is not allowed.
Position of modules and IP port connections
The positions of the modules are important to address them by IP ports,
which is required by the chromatography software PurityChrom® and by
single device configuration.
One module
Is one module in the assistant, it is always connected to the left (port
10002).
Two modules
For two modules in the assistant, port 10003 is not connected.
Fig. 4
ASM 2.1L gas outlet on the front