ASM 2.1L user manual V6815
11
Safety for users
Safety for users
Professional Group
The user manual addresses persons who are qualified as chemical labora-
tory technicians or have completed comparable vocational training.
The following knowledge is required:
Fundamental knowledge of liquid chromatography
Knowledge regarding substances that are suitable only to a limited
extent for use in liquid chromatography
Knowledge regarding the health risks of chemicals
Participation during an installation of a device or a training by the com-
pany KNAUER or an authorized company.
If you do not belong to this or a comparable professional group, you may
not perform the work described in this user manual under any circum-
stances. In this case, please contact your superior.
Safety Equipment
When working with the device, take measures according to lab regula-
tions and wear protective clothing:
Safety glasses with side protection
Protective gloves
Lab coat
What must be taken into
account?
All safety instructions in the user manual
The environmental, installation, and connection specifications in the
user manual
National and international regulations pertaining to laboratory work
Original spare parts, tools, and solvents made or recommended by
KNAUER
Good Laboratory Practice (GLP)
Accident prevention regulations published by the accident insurance
companies for laboratory work
Filtration of substances under analysis
Use of inline filters
Once they have been used, never re-use capillaries in other areas of
the HPLC system.
Only use a given PEEK fitting for one specific port and never re-use it
for other ports. Always install new PEEK fittings on each separate port.
Follow KNAUER or manufacturer’s instructions on caring for the
colums.
More safety-relevant information is listed below:
flammability: Organic solvents are highly flammable. Since capillaries
can detach from their screw fittings and allow solvent to escape, it is
prohibited to have any open flames near the analytical system.
solvent tray: Risk of electrical shock or short circuit if liquids get into the
device's interior. For this reason, place all bottles in a solvent tray.
solvent lines: Install capillaries and tubing in such a way that liquids
cannot get into the interior in case of a leak.
leaks: Regularly check if any system components are leaking.
power cable: Defective power cables are not to be used to connect the
device and the power supply system.
self-ignition point: Only use eluents that have a self-ignition point
higher than 150 °C under normal ambient conditions.
Installation Qualification Document (“IQ”, English)