ASM 2.1L user manual V6815
Technical data
28
Eye injury
Irritation of retina through UV light. Concentrated UV light can leak out
from the flow cell or the fiber optic connectors.
Switch off the device and pull the power plug.
Technical data
Communication
General
Repeat orders
This list for repeat orders is valid for the time the document has been pub-
lished. Deviations afterwards are possible.
Note:
For repeat orders of spare parts use the enclosed packing list. Con-
tact the Technical Support in case there are any questions on spare parts
or accessories.
Further Information
Further information on spare parts and accessories can be found online:
www.knauer.net
Module
Measures
> 2000 operating
hours
Detector
Check light intensity of the D2 lamp
Inspect the flow-cell assembly.
Clean the flow cell.
Replace the flow cell.
Pump
Control the seals.
Valve
Check screw fittings for leak loss.
Degasser
Check screw fittings for leak loss.
> 5000 operating
hours
Pump
Replace the pistons
Replace the check valves
Valve
Check screw fittings for leak loss.
> 10000 operating
hours
Pump
Replace the O-rings at the plates
Operation
Digital: via LAN, Mobile Control
Analog: Integrator (±2.5 V), Pin header
Power supply
100–240 V; 50– 60 Hz; maximum 100 W
Dimensions
361 × 158 × 523 mm (Width × Height ×
Depth)
Dimensions
361 × 158 × 523 mm (Width × Height ×
Depth)
Leak sensor
yes
Ambient conditions
Temperature range: 10–40 °C;
50–104 °F
Air humidity: 10–90 %, non-condensing