![KLS Martin marLED X Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/kls-martin/marled-x/marled-x_service-manual_1978990084.webp)
84
9.4
marTouch® Pro on the Light Head
1 Connecting cable JST 5-pole
2 USB port for firmware update
3 Connection cable on the side of the light
Assignment JST 5-pole
Pin
Signal
Cable color
1
mLB L
GY
2
mLB H
YE
3
mLB GND
GN
4
VDC-
BN
5
VDC+
WH
In the event of an error, measure voltages on the
cable (3) between:
•
mLB L and mLB GND: 1.5–3 VDC
•
mLB H and mLB GND: 1.5–3 VDC