Test Instructions
If
N/A
has been ticked in at least one position even though the object to be tested is present, it is
mandatory to indicate in the field
Comment
the reason why it was not possible to carry out the
corresponding test (e.g. object not accessible, defective).
1. Checking the wall/ceiling assembly
a) Checking the wall/ceiling assembly for stability
Check the ceiling assembly for stability!
According to the manufacturer, heavy-duty anchors or expansion dowels may be installed only once.
During inspection, the tightening torques thus cannot be re-checked using a torque wrench.
•
During an inspection, check visually and by mechanical testing as described in 1.b).
If the screws of the ceiling assembly are not accessible (e.g. closed air cover), the suspension arm system
and the ceiling assembly can be checked using the following test method.
•
Check the stability of the wall/ceiling assembly by appropriate shaking of the tracking arm
over the entire rotational range of the operating light: