15
Recycling
KITETT® breast pumps must be disposed of in accordance with local disposal guidelines for electrical
appliances. Use your local waste disposal system for electrical and electronic equipment (including batteries).
You thus contribute to the preservation of the environment and the protection of human health.
Device placed on the market after 13 August 2005.
10.
Technical specifications
Device class (in accordance with Directive 93/42/EEC
amended by Directive 2007/47/EC)
IIa
Class of electrical safety device
Class II
Type of applied parts
BF
Degree of protection
IP22
Expected life of accessories
1 year
Operating life span of the unit
300 hours
Useful life span of the unit
3 years
Capacity of the collection container
100ml
Weight of the device
200g without battery
Dimensions
120mm x 90mm x 60mm
Voltage
Input: 100-240V (AC)
–
50/60Hz
–
0.5A Max. /
Output: 6V (DC)
–
1.0A
Length of the cable
1.8m
Absorbed power
5.2W
Suction Values
Stimulation phase: -80 to -140 mbar
Expression phase: -155 to -280 mbar
Maximum suction: <-330 mbar
Rate
40 to 120 cycles/min.
Sound level
< 56 dB(A)
Power supply
Alternating current 100-240V - 50/60 Hz
Battery
4 x 1,5V - AA, LR06
Room temperature (T)
for operation
+5°C < T < +40°C
Room temperature (T)
for transport and storage
-25°C < T < +70°C
Relative humidity (H)
for transport, storage and operation
15% < H < 90%
Atmospheric pressure (P)
for transport, storage and operation
700 hPa < P < 1060 hPa
Essential performance features:
•
Provide adequate suction so that mothers can express their milk.
•
Control the rate of suction to best meet
the baby’s sucking rhythm.
•
Equipment with moderate compression and low flow. Intermittent suction.
Part applied to the body:
breast shield
NOTE:
The CE declaration of conformity and a technical description can be supplied on demand.
Identification:
Each breast pump has a batch identification that indicates the year of manufacture.
11.
Electromagnetic compatibility (EMC)
Electromedical devices require special precautions with regard to EMC and must be installed and set to operate in
accordance with the EMC information provided in this manual.
WARNING:
Avoid using this unit near or stacked onto other equipment as this may cause a malfunction. If this is
necessary, this appliance and the other appliances should be monitored to check they are operating normally.
Содержание MINIKIT DUO
Страница 2: ...2...
Страница 3: ...3...
Страница 4: ...4...
Страница 21: ...21 S L 3 a b 1 KITETT KOLOR 2 Clean Valve 3 4 5 20 C 1 Clean Valve 2 3 3...
Страница 22: ...22 c 1 KITETT KOLOR 2 Clean Valve 3 4 5 1 1 10 1000 2 3 d KITETT KOLOR Clean Valve 20 C 10 4 a MINIKIT DUO 1...
Страница 24: ...24 c 1 2 3 4 5 MINIKIT DUO 1 4 2 1 2 45 2 3 90 3 4 135 4 5 180 5 5 9 6...
Страница 25: ...25 MINIKIT DUO 30 d 1 2 Clean Valve 3 4 MINIKIT DUO 5 6 3 5 a www dtf fr b MINIKIT DUO 2...
Страница 26: ...26 6 3000 DTF MEDICAL 10 1 42 5 C...
Страница 32: ...32 13 2005 60601 1 II 13 2005...
Страница 35: ...35 b MINIKIT DUO KIT EXPRESSION KOLOR MINIKIT DUO KOLOR 4 2 4 4 2 S 76 L 90 Kolorama KOLORYOU KOLOR 2 S L...
Страница 36: ...36 3 a b 1 2 3 4 5 20 C 1 2 3 3 c 1 2 3 4 5...
Страница 37: ...37 1 1 10 1000 2 3 d Clean Valve 20 C 10 4 a MINIKIT DUO 1 30 30...
Страница 39: ...39 c 1 2 3 4 5 MINIKIT DUO 1 4 2 1 45 2 2 90 3 3 135 4 4 180 5 5 5 9 6 30 MINIKIT DUO...
Страница 40: ...40 d 1 2 Clean Valve 3 4 MINIKIT DUO 5 2 6 3 5 a www dtf fr b MINIKIT DUO 2 6 3000...
Страница 41: ...41 DTF MEDICAL 10 1 42 5 C 7 DTF MEDICAL MINIKIT DUO Clean Valve DTF MEDICAL...
Страница 46: ...46 2005 13 IEC 60601 1 2005 13...
Страница 49: ...49 b MINIKIT DUO KOLOR MINIKIT DUO KOLOR 4 2 4 4 2 S 76 L 90 Kolorama KOLORYOU KOLOR 2...
Страница 50: ...50 S L 3 a b 1 2 3 4 5 20 C 1 2 3 3...
Страница 51: ...51 c 1 2 3 4 5 10 1000 2 3 d Clean Valve 20 C 10 4 a MINIKIT DUO 1...
Страница 53: ...53 6 c 1 2 3 4 5 MINIKIT DUO 1 4 2 1 2 45 2 3 90 3 4 135 4 5 180 5 5 9 6...
Страница 54: ...54 MINIKIT DUO 30 d 1 2 Clean Valve 3 4 MINIKIT DUO 5 2 6 3 5 a 25 C 4 C 18 C 6 8 12 4 24...
Страница 56: ...56 10 1 42 5 7 DTF MEDICAL Clean Valve MINIKIT DUO DTF MEDICAL DTF MEDICAL...
Страница 61: ...61 13 2005 IEC 60601 1 II 13 2005...
Страница 76: ...76...
Страница 77: ...77...