4. Set the different parameters of the analyser

The 

“Settings”

 menu allows to set the following parameters : 

Instrument (screen, printing, auto-extinction,...)

Combustibles

Measurement

References (O

2

, atmospheric pressure and altitude)

AutoZ/Purge

Operators

The “Settings” part allows to set parameters that can affect the performed measurements. These settings must be made by a 
qualified technician.

4.1. Set the instrument 

4.1.1. Screen setting

This part allows to set brightness, contrast and backlight of the screen from 1 to 9. 

Turn on the instrument.

Home screen is displayed with measurement menu at the forefront.

Go to 

“Settings” 

with rotative button then press 

OK

.

Go to 

“Instrument”

 then press 

OK

.

The screen with the different parameters to set (brightness, contrast and backlight) is displayed.

Press 

OK

.

 

The screen with the settings to perform appears.

Go to the set to perform with the rotative button then press 

OK

.

The name of the setting appears in black on blue background.

Turn the rotative button on the left to decrease 

brightness, contrast or backlight

 or turn the rotative button on the right to increase it.

4.1.2. Set the auto-extinction

The auto-extinction turn off the analyser automatically after some time of non-use.
It is possible to set the auto-extinction at 15 / 30 / 45 / 60 / 90 or 105 minutes or to deactivate it.

“Settings instrument”

 screen is displayed.

Go to 

“Auto-extinction”

 with the rotative button then press 

OK

Select the required time of auto-extinction or 

“Off”

 to deactivate it with the rotative button.

Press 

OK

.

4.1.3. Change the language

It is possible to select the language of the instrument : French or English

“Settings instrument”

 screen is displayed.

Go to 

“Language”

 with the rotative button then press 

OK

Select the language with the rotative button.

Press 

OK

.

4.1.4. Modify date and time

“Settings instrument”

 screen is displayed.

Go to 

“Date/time”

 with the rotative button then press 

OK

.

Go to the parameter to modify with the rotative button (day, month, year, hour, minute or second).

Press 

OK

.

Turn the rotative button to modify the selected parameter.

Press 

OK

.

Repeat the procedure on another parameter if required.

Go to 

“Date Format”

 with the rotative button then press 

OK

.

Select 

« Day/Month »

 to display the day then the month or 

“Month/Day”

 to display the month then the day then press 

OK.

Go to 

“Time Format”

 with the rotative button then press 

OK

.

Select the time format : 

“12” 

or

 “24”.

If12 is selected, 

“AM”

 for ante meridiem or 

« PM » 

for

 

post meridiem will be displayed in front of  the time.

Press the 

“Validate” 

function key to quit the screen and save the modifications or press 

Esc

 to cancel.

4.1.5. Ticket printing setting

“Settings instrument”

 screen is displayed.

Go to 

“Printing”

 with the rotative button then press 

OK

.

Set the different parameters of the analyser

15

netzerotools.com

netzerotools.com

Содержание Kigaz 300 CLA

Страница 1: ...netzerotools com netzerotools com netzerotools com Kimo Kigaz 300 CLA Combustion Gas Analyzer Kimo Kigaz 300 STD Combustion Gas Analyser Kimo Kigaz 300 PRO Combustion Gas Analyzer...

Страница 2: ...netzerotools com netzerotools com...

Страница 3: ...s 17 4 3 2 Set measurement units 18 4 3 3 Set the CO protection threshold 18 4 4 Set the references 18 4 5 Set the duration of autozero and purge time 18 4 6 Set the operators 18 4 6 1 Add an operator...

Страница 4: ...rform the measurement with the probe 32 17 2 Perform the measurement with the sensor 32 17 3 Link the measurement to a customer 32 17 4 Print the measurement on the ticket 32 18 Perform a measurement...

Страница 5: ...he instrument 48 a The Internet connection is not protected by a proxy 48 b The Internet connection is protected by a proxy 48 6 8 3 Install a 3rd language 49 6 8 4 Send a report of the device to the...

Страница 6: ...netzerotools com netzerotools com...

Страница 7: ...in ABS or metallic optional with a magnetic protective cover a differential pressure kit an interchangeable flue gas probe of 300 mm length with integrated draft and LED a Li Ion battery and its char...

Страница 8: ...temperature Pt100 CH4 Flue gas probe connection P pressure tap P pressure tap Power supply USB connection Top view Bottom view Side view Function keys On Off button OK button Esc button Rotating butto...

Страница 9: ...NTC thermoresistance for the measurement of internal temperature this sensor is also used for the measurement of ambient temperature If a temperature control of the combustive air must be performed di...

Страница 10: ...gases SO2 concentration of SO2 in flue gases Tf flue gases temperature Ta combustive air temperature Calculated values Air Excess connection between the volume of combustive air and the requested vol...

Страница 11: ...in external probe From 50 to 250 C 0 1 C 0 3 of measured value 0 25 C 30 s Dewpoint temperature Calculated From 0 to 99 Ctd 0 1 C DHW temperature TcK external probe From 200 to 1300 C 0 1 C 1 C Draft...

Страница 12: ...e probe and temperature sensor could be destroyed After measurements before packing the analyser wait for the probe to be cold As long as it is hot the probe could produce injuries The exhalations of...

Страница 13: ...ome screen is displayed with measurement menu at the forefront Insert the probe in the flue as show on the figure for at least 3 minutes Press OK The screen with the different possible measurements is...

Страница 14: ...with the rotative button then press OK or create a customer see page 20 then select it The analyser displays the list of available boilers Select the boiler with the rotative button then press OK or c...

Страница 15: ...or to deactivate it Settings instrument screen is displayed Go to Auto extinction with the rotative button then press OK Select the required time of auto extinction or Off to deactivate it with the ro...

Страница 16: ...e with the rotative button Press OK to activate it ON or to deactivate it OFF 4 1 8 Activate or deactivate the vocal support The vocal support allows to activate a voice that gives the details during...

Страница 17: ...en backs to the list of combustible The new combustible in well on the list in last position Every coefficient of the new combustible are set at 0 Follow the procedure of modification of combustible c...

Страница 18: ...the threshold Press Esc to quit the screen 4 4 Set the references Settings References screen allows to set the features of the measurement environment O2 reference altitude and atmospheric pressure Th...

Страница 19: ...the function key A message is displayed and asks to confirm the return to factory settings Go to OK with the rotative button then press OK To cancel this return to factory settings when the Back to fa...

Страница 20: ...both following lines of the address if necessary Go to Tel line with the rotative button then press OK Enter customer s telephone number with the rotative button and press OK to validate the letters P...

Страница 21: ...e boiler then press OK Press OK to tick the Air pulser burner box if the boiler has one Press the function key to save the boiler features The analyser backs to the list of boilers 6 2 Modify a boiler...

Страница 22: ...the required customer then press OK Every boilers linked to this customer are displayed Go to the required boiler then press OK Every inspections linked to this customer and boiler are displayed Go to...

Страница 23: ...Go to Customized with the rotative button then press OK The list of procedures saved in the analyser is displayed Select the procedure with the rotative button then press OK The analyser displays the...

Страница 24: ...mbar Butane Propane NF Pu KW x 2 166 Pu KW x 2 519 Pu KW x 0 651 Pu KW x 0 843 CE Pu KW x 1 927 Pu KW x 2 241 Pu KW x 0 579 Pu KW x 0 750 9 2 Perform a measurement of real flow The measurement of the...

Страница 25: ...bar then press OK A message indicating Please wait 10 minutes and a progress bar are displayed At the end of the 10 minutes results are displayed start pressure end pressure possible loss of pressure...

Страница 26: ...e The analyser displays the list of registered customers Select a customer with the rotative button then press OK or create a new customer see page 20 then select it The analyser displays the list of...

Страница 27: ...tomer registered in the analyser or to a customer unregistered in the analyser that must be created Press the function key to save The analyser displays the list of registered customers Select a custo...

Страница 28: ...een is displayed Press OK Use the rotative button to select the required measurement duration 30 40 50 s or 1 2 3 4 5 minutes Press OK to validate the measurement duration Press Esc to back to the scr...

Страница 29: ...perature menu during the autozero Go to Internal Temp with the rotative button then press OK The screen displays the NTC temperature detected by the analyser 14 4 Perform a dewpoint temperature measur...

Страница 30: ...zero remaining time and a progress bar display on the screen However it is possible to reach the Bacharach index menu during the autozero Press OK Go to Bacharach index with the rotative button Press...

Страница 31: ...n the boiler The measured ionisation current is displayed on the analyser This value must be equal to several dozen of A 16 2 Link the measurement to a customer When the measurement has been performed...

Страница 32: ...to Measurement menu with the rotative button then press OK The analyser performs the autozero remaining time and a progress bar display on the screen Go to Combustion with the rotative button then pre...

Страница 33: ...e menu then press OK Go to Draught Press then press OK The analyser displays the measured values 18 1 Link the measurement to a customer When the measurement has been performed it is possible to link...

Страница 34: ...analyser type of sensor its serial number and its installation date To get to this menu From the home screen Go to Information menu with the rotative button then press OK The information is displayed...

Страница 35: ...ars Annual CO H2 From 2 to 3 years Annual NO From 2 to 3 years Annual NO2 From 2 to 3 years Annual SO2 From 2 to 3 years Annual CH4 From 3 to 5 years Annual 20 2 Replace the sensors Follow this proced...

Страница 36: ...rs NO 48 hours NO2 48 hours SO2 48 hours CH4 48 hours 20 3 Replace the paper roll of the printer Follow this procedure to replace the paper roll of the printer Raise the upper part of the printer Put...

Страница 37: ...filter When the filter inside the analyser probe is too much clogged black coloured it is necessary to change it to get accurate measurement and to avoid to damage the instrument and its sensors Rotat...

Страница 38: ...r this value has to be entered in the device in O2 oxygen value measured in VF the theoretical volume of flue gas in m3 given in the table of coefficient NCV the net calorific value in MJ given in the...

Страница 39: ...Calculations of the different parameters 39 netzerotools com netzerotools com...

Страница 40: ...m configuration For the proper functioning of the software the following minimum configuration is recommended Operating system Windows XP VISTA 7 Communication port USB 2 0 DVD player RAM memory 1 GO...

Страница 41: ...rs database window Database settings opens the saving settings window of the database Connection state KIGAZ Unload KIGAZ allows to unload data saved in the analyser and the instrument setting Setting...

Страница 42: ...vice About indicates software version and database version 3 2 Meaning and functions of tool bar buttons 2 Closes all the opened windows Opens the boilers database window Opens the total database wind...

Страница 43: ...hen on Unload kigaz The software recovers inspections contained in the analyser This operation can take a few minutes At the end of the recovery the following window is displayed Tick the correspondin...

Страница 44: ...e menus bar then click on Database settings The opposite window opens Click on Browse to define the location where the databases will be saved Tick then select the repetition of saving for a recurring...

Страница 45: ...and Browse button is active Click on Browse button in front Customer to link the boiler to a customer Fill in Model and Brand fields compulsory fields Click on Browse button in front of Combustible an...

Страница 46: ...tition it means the type of automatic saving of the database Select the saving frequency of the database between 1 hour and 30 hours or between 1 day and 30 days OR Tick When closing software to autom...

Страница 47: ...ton to send modifications to the analyser 6 2 Change the references This window allows to modify the O2 reference oxygen level in which the analyser is used altitude altitude in which the analyser is...

Страница 48: ...m right of the window Click on Get the last firmware button The opposite window opens Two choices are then possible a The Internet connection is not protected by a proxy Click on Cancel button on the...

Страница 49: ...k if the update has been taken into account 6 8 3 Install a 3rd language Select the language to send in the drop down list box Click on Send 3rd language button 6 8 4 Send a report of the device to th...

Страница 50: ...durand orange fr Mail server server address used to send email ex smtp orange fr Mail subject Mail format plain text or html Advanced options by clicking on this the email setting window opens and al...

Страница 51: ...netzerotools com netzerotools com...

Страница 52: ...netzerotools com netzerotools com...

Отзывы: