DME1600
31
Trigger Commands
Messa
ge Ref
e
re
nce
2
TRIG:SOUR
Sets the trigger source. The DME1600 can accept a software (BUS) trigger, an immediate
internal trigger, or a hardware trigger applied to the EXT TRIG input terminal on the rear
panel. The default setting is for the DME1600 to use an immediate trigger.
Command
TRIGger:SOURce {BUS|IMMediate|EXTernal}
TRIGger:SOURce?
Response
In response to TRIG:SOUR?, the DME1600 returns the trigger source.
TRIG:DEL
Sets the trigger delay time in seconds. The trigger delay is the time between a trigger signal
and the measurements (samples) that follow it. You can set the delay to a value between 0
and 3600 seconds.
Command
TRIGger:DELay {<seconds>|MINimum|MAXimum}
TRIGger:DELay?
Response
In response to TRIG:DEL?, the DME1600 returns the trigger delay.
TRIG:DEL:AUTO
Turns the auto trigger delay on or off. The delay is determined according to the measurement
function, range, integration time, and AC filter settings. If you specify the delay time, the auto
trigger delay is automatically disabled.
Command
TRIGger:DELay:AUTO {OFF|ON}
TRIGger:DELay:AUTO?
Response
In response to TRIG:DEL:AUTO?, the DME1600 returns “0” (OFF) or “1” (ON).
SAMP:COUN
Sets the number of measurements (samples) per trigger. You can set the number of mea-
surements per trigger to a number from 1 to 50000.
Command
SAMPle:COUNt {<value>|MINimum|MAXimum}
SAMPle:COUNt? [MINimum|MAXimum]
Response
In response to SAMP:COUN?, the DME1600 returns the number of measured values per
trigger.
TRIG:COUN
Sets the number of triggers that the DME1600 receives before it returns to the idle state to a
number between 1 and 50000. You can specify the INFinite parameter to make the
DME1600 receive triggers continuously. The trigger count is ignored in local mode.
Command
TRIGger:COUNt {<value>|MINimum|MAXimum|INFinite}
TRIGger:COUNt? [MINimum|MAXimum|INFinite]
Response
In response to TRIG:COUN, the DME1600 returns the trigger count. If INFinite is selected,
the DME1600 returns “9.9037.”
Содержание DME1600
Страница 4: ...4 DME1600 This page left blank intentionally...
Страница 5: ...Setup This chapter explains the settings that you need to configure to use the interfaces 1...
Страница 9: ...Message Reference This chapter explains the SCPI com mands 2 Message Reference...
Страница 43: ...Appendix A Error Messages B Sample Programs...
Страница 55: ......