Errors are detected and saved to the error queue in first in, first out (FIFO) order. The error
that was detected first is displayed first. When all the messages in the error queue are
retrieved, the ERROR indicator turns off. The DME1600 beeps once each time an error
occurs.
When more than 20 errors have been detected, the last error stored in the error queue (the
most recent error) changes to “-350, Too many error.” As long as no errors are deleted from
the queue, no additional errors can be stored, and the DME1600 will show “+0, No error.”
The error queue is cleared when the power is turned off or a *CLS (clear status) command is
executed. The *RST (reset) command does not clear the error queue.
Error code
Description
-101 Invalid character
An invalid character was found in the command string.
-102 Syntax error
Invalid syntax was found in the command string.
-103 Invalid separator
An invalid separator was found in the command string.
-104 Data type error
A parameter type error was found in the command string.
-105 GET not allowed
A Group Execute Trigger (GET) is not allowed in the command string.
-108 Parameter not allowed
More parameters were found than needed for the command.
-109 Missing parameter
Not enough parameters were received for the command.
-112 Program mnemonic too long
A command header with too many characters was received.
-113 Undefined header
An invalid command was received.
-121 Invalid character in number
An invalid character was found in the number specified for a parameter
value.
-123 Numeric overflow
A numeric parameter with exponent larger than 32000 was found.
-124 Too many digits
A numeric parameter was found whose mantissa contained more than 255
digits, excluding leading zeros.
-131 Invalid suffix
A suffix was incorrectly specified for a numeric parameter.
-138 Suffix not allowed
A suffix was received following a numeric parameter which does not accept
a suffix.
-148 Character not allowed
A discrete parameter was received but a character string or a numeric
parameter was expected.
-151 Invalid string data
An invalid character string was received.
-158 String data not allowed
A character string was received but not allowed for the command.
-160 to -168 Block data errors
Block data is not acceptable.
-170 to -178 Expression errors
The meter does not accept mathematical expression.
-211 Trigger ignored
A Group Execute Trigger (GET) or *TRG was received but the trigger was
ignored.
-213 Trigger deadlock
A trigger deadlock occurs when the trigger source is BUS and a READ?
Command is received.
-214 Init Ignored
An INITiate command was received but could not be executed because a
measurement was already in progress. Send a device clear to halt a mea-
surement in progress and place the meter in the “idle” state.
-221 Settings conflict
This error can be generated in one of the following situations:
Situation 1: You sent a CONFigure or MEASure command with autorange enabled and
with a fixed resolution.
Situation 2: Situation 2: You turned math on and then changed to a math operation that
was not valid with the present measurement function.
-222 Data out of range
A numeric parameter value is out of range.
Содержание DME1600
Страница 4: ...4 DME1600 This page left blank intentionally...
Страница 5: ...Setup This chapter explains the settings that you need to configure to use the interfaces 1...
Страница 9: ...Message Reference This chapter explains the SCPI com mands 2 Message Reference...
Страница 43: ...Appendix A Error Messages B Sample Programs...
Страница 55: ......