DME1600
Measurement Configuration Commands
AVER:COUN
Sets the number of digital filter stack entries (2 to 100).
Command
[SENSe:]AVERage:COUNt {<value>|MINimum|MAXimum}
[SENSe:]AVERage:COUNt? [MINimum|MAXimum]
Response
In response to AVER:COUN?, the DME1600 returns the number of digital filter stack entries.
AVER:STAT
Turns the digital filter on or off. The default setting is “ON.”
Command
[SENSe:]AVERage:STATe {OFF|ON}
[SENSe:]AVERage:STATe?
Response
In response to AVER:STAT?, the DME1600 returns “0” (OFF) or “1” (ON).
ZERO:AUTO
Turns auto zero adjustment on or off. The default setting is “ON.” The results for OFF and
ONCE are similar. When auto zero adjustment is set to OFF, the DME1600 does not perform
a new offset measurement until it is in a trigger-wait state. When auto zero adjustment is set
to ONCE, an offset measurement is performed immediately.
Command
[SENSe:]ZERO:AUTO {OFF|ONCE|ON}
[SENSe:]ZERO:AUTO?
Response
In response to ZERO:AUTO?, the DME1600 returns “1” (ON) or “0” (OFF or ONCE).
GAIN:AUTO
Turns auto gain on or off. The default setting is “ON.” The results for OFF and ONCE are sim-
ilar. When auto gain is set to OFF, the DME1600 does not perform a new offset measurement
until it is in a trigger-wait state. When auto zero adjustment is set to ONCE, an offset mea-
surement is performed immediately.
Command
[SENSe:]GAIN:AUTO {OFF|ONCE|ON}
[SENSe:]GAIN:AUTO?
Response
In response to GAIN:AUTO?, the DME1600 returns “1” (ON) or “0” (OFF or ONCE).
INP:IMP:AUTO
Turns auto input resistance mode for DC voltage measurement on or off. The default setting
is “OFF.” When auto input resistance mode is set to ON, the input resistance for the 100 mV,
1 V, and 10 V ranges is set to > 10 G
. When auto input resistance mode is set to OFF, the
input resistance for all ranges is fixed at 10 M
.
Command
INPut:IMPedance:AUTO {OFF|ON}
INPut:IMPedance:AUTO?
Response
In response to INP:IMP:AUTO?, the DME1600 returns “1” (ON) or “0” (OFF).
Содержание DME1600
Страница 4: ...4 DME1600 This page left blank intentionally...
Страница 5: ...Setup This chapter explains the settings that you need to configure to use the interfaces 1...
Страница 9: ...Message Reference This chapter explains the SCPI com mands 2 Message Reference...
Страница 43: ...Appendix A Error Messages B Sample Programs...
Страница 55: ......