Measurement Configuration Commands
Messa
ge Ref
e
re
nce
2
<function>:RANGe:AUTO
Turns the auto range feature for the selected measurement function on or off. The default set-
ting is “ON.” For frequency or period measurement, use FREQuency:VOLTage or
PERiod:VOLTage. When the auto range feature is enabled, the next lower range is selected
when the signal falls below 10 % of the current range, and the next higher range is selected
when the input signal exceeds 120 % of the current range.
Use one of the following character strings to indicate the function name (<function>).
Command
[SENSe:]<function>:RANGe:AUTO {OFF|ON}
[SENSe:]<function>:RANGe:AUTO?
Response
In response to <function>:RANG:AUTO?, the DME1600 returns “1” (ON) or “0” (OFF).
<function>:RES
Sets the resolution of the selected measurement function (this command does not apply to
the frequency, period, or ratio measurement functions). Specify the resolution using the same
unit as the measurement function, not using the number of digits. MIN is the minimum allow-
able value (the maximum value for the resolution). MAX is the maximum allowable value (the
minimum value for the resolution).
Use one of the following character strings to indicate the function name (<function>).
Command
[SENSe:]<function>:RESolution {<resolution>|MINimum|MAXimum}
[SENSe:]<function>:RESolution? [MINimum|MAXimum]
Response
In response to <function>:RES?, the DME1600 returns the resolution of the selected
measurement function.
UNIT
Sets the unit used for temperature measurement. Cel indicates Celsius, Far indicates Fahr-
enheit, and K indicates Kelvin. The default value is “Cel.”
Command
[SENSe:]UNIT {Cel|Far|K}
[SENSe:]UNIT?
Response
In response to UNIT?, the DME1600 returns the unit used for temperature measurement.
TCO:TYPE
Sets the thermocouple type. The default setting is “K.”
Command
[SENSe:]TCOuple:TYPE {E|J|K|N|R|S|T}
[SENSe:]TCOuple:TYPE?
Response
In response to TCO:TYPE?, the DME1600 returns the thermocouple type.
VOLTage:DC
VOLTage:AC
CURRent:DC
CURRent:AC
RESistance (2-wire resistance) FRESistance (4-wire resistance)
FREQuency:VOLtage
PERiod:VOLtage
VOLTage:DC
VOLTage:AC
CURRent:DC
CURRent:AC
RESistance (2-wire resistance)
FRESistance (4-wire resistance)
Содержание DME1600
Страница 4: ...4 DME1600 This page left blank intentionally...
Страница 5: ...Setup This chapter explains the settings that you need to configure to use the interfaces 1...
Страница 9: ...Message Reference This chapter explains the SCPI com mands 2 Message Reference...
Страница 43: ...Appendix A Error Messages B Sample Programs...
Страница 55: ......